PMID- 11213924 OWN - NLM STAT- MEDLINE DCOM- 20010419 LR - 20170214 IS - 0094-2405 (Print) IS - 0094-2405 (Linking) VI - 28 IP - 1 DP - 2001 Jan TI - Leaf position error during conformal dynamic arc and intensity modulated arc treatments. PG - 67-72 AB - Conformal dynamic arc (CD-ARC) and intensity modulated arc treatments (IMAT) are both treatment modalities where the multileaf collimator (MLC) can change leaf position dynamically during gantry rotation. These treatment techniques can be used to generate complex isodose distributions, similar to those used in fix-gantry intensity modulation. However, a beam-hold delay cannot be used during CD-ARC or IMAT treatments to reduce spatial error. Consequently, a certain amount of leaf position error will have to be accepted in order to make the treatment deliverable. Measurements of leaf position accuracy were taken with leaf velocities ranging from 0.3 to 3.0 cm/s. The average and maximum leaf position errors were measured, and a least-squares linear regression analysis was performed on the measured data to determine the MLC velocity error coefficient. The average position errors range from 0.03 to 0.21 cm, with the largest deviations occurring at the maximum achievable leaf velocity (3.0 cm/s). The measured MLC velocity error coefficient was 0.0674 s for a collimator rotation of 0 degrees and 0.0681 s for a collimator rotation of 90 degrees. The distribution in leaf position error between the 0 degrees and 90 degrees collimator rotations was within statistical uncertainty. A simple formula was developed based on these results for estimating the velocity-dependent dosimetric error. Using this technique, a dosimetric error index for plan evaluation can be calculated from the treatment time and the dynamic MLC leaf controller file. FAU - Ramsey, C R AU - Ramsey CR AD - Department of Radiation Oncology, Thompson Cancer Survival Center, Knoxville, Tennessee 37921, USA. cramsey@utk.edu FAU - Spencer, K M AU - Spencer KM FAU - Alhakeem, R AU - Alhakeem R FAU - Oliver, A L AU - Oliver AL LA - eng PT - Journal Article PT - Research Support, Non-U.S. Gov't PL - United States TA - Med Phys JT - Medical physics JID - 0425746 SB - IM MH - Biophysical Phenomena MH - Biophysics MH - Humans MH - Male MH - Prostatic Neoplasms/radiotherapy MH - Radiotherapy Dosage MH - Radiotherapy, Conformal/instrumentation/*methods/statistics & numerical data EDAT- 2001/02/24 12:00 MHDA- 2001/04/21 10:01 CRDT- 2001/02/24 12:00 PHST- 2001/02/24 12:00 [pubmed] PHST- 2001/04/21 10:01 [medline] PHST- 2001/02/24 12:00 [entrez] AID - 10.1118/1.1333410 [doi] PST - ppublish SO - Med Phys. 2001 Jan;28(1):67-72. doi: 10.1118/1.1333410.