PMID- 11747997 OWN - NLM STAT- MEDLINE DCOM- 20020307 LR - 20191105 IS - 0958-3947 (Print) IS - 1873-4022 (Linking) VI - 26 IP - 4 DP - 2001 Winter TI - Quantitative analysis of errors in fractionated stereotactic radiotherapy. PG - 315-8 AB - Fractionated stereotactic radiotherapy (FSRT) offers a technique to minimize the absorbed dose to normal tissues; therefore, quality assurance is essential for these procedures. In this study, quality assurance for FSRT of 58 cases, between August 1995 and August 1997 are described, and the errors for each step and overall accuracy were estimated. Some of the important items for FSRT procedures are: accuracy in CT localization, transferred image distortion, laser alignment, isocentric accuracy of linear accelerator, head frame movement, portal verification, and various human errors. A geometric phantom, that has known coordinates was used to estimate the accuracy of CT localization. A treatment planning computer was used for checking the transferred image distortion. The mechanical isocenter standard (MIS), rectilinear phantom pointer: (RLPP), and laser target localizer frame (LTLF) were used for laser alignment and target coordinates setting. Head-frame stability check was performed by a depth confirmation helmet (DCH). A film test was done to check isocentric accuracy and portal verification. All measured data for the 58 patients were recorded and analyzed for each item. 4-MV x-rays from a linear accelerator, were used for FSRT, along with homemade circular cones with diameters from 20 to 70 mm (interval: 5 mm). The accuracy in CT localization was 1.2+/-0.5 mm. The isocentric accuracy of the linear accelerator, including laser alignment, was 0.5+/-0.2 mm. The reproducibility of the head frame was 1.1+/-0.6 mm. The overall accuracy was 1.7+/-0.7 mm, excluding human errors. FAU - Choi, D R AU - Choi DR AD - Department of Medical Physics, Tom Baker Cancer Centre, University of Calgary, Alberta, Canada. FAU - Kim, D Y AU - Kim DY FAU - Ahn, Y C AU - Ahn YC FAU - Huh, S J AU - Huh SJ FAU - Yeo, I J AU - Yeo IJ FAU - Nam, D H AU - Nam DH FAU - Lee, J I AU - Lee JI FAU - Park, K AU - Park K FAU - Kim, J H AU - Kim JH LA - eng PT - Journal Article PT - Research Support, Non-U.S. Gov't PL - United States TA - Med Dosim JT - Medical dosimetry : official journal of the American Association of Medical Dosimetrists JID - 8908862 SB - IM MH - Brain Neoplasms/diagnostic imaging/*radiotherapy MH - *Dose Fractionation, Radiation MH - *Evaluation Studies as Topic MH - Humans MH - Image Processing, Computer-Assisted MH - Mathematical Computing MH - *Medical Errors MH - Particle Accelerators MH - *Quality Assurance, Health Care MH - Radiotherapy Planning, Computer-Assisted MH - Reproducibility of Results MH - *Stereotaxic Techniques MH - Tomography, X-Ray Computed EDAT- 2001/12/19 10:00 MHDA- 2002/03/08 10:01 CRDT- 2001/12/19 10:00 PHST- 2001/12/19 10:00 [pubmed] PHST- 2002/03/08 10:01 [medline] PHST- 2001/12/19 10:00 [entrez] AID - S0958394701000826 [pii] AID - 10.1016/s0958-3947(01)00082-6 [doi] PST - ppublish SO - Med Dosim. 2001 Winter;26(4):315-8. doi: 10.1016/s0958-3947(01)00082-6.