PMID- 12707697 OWN - NLM STAT- MEDLINE DCOM- 20031112 LR - 20131121 IS - 0085-4530 (Print) IS - 0085-4530 (Linking) VI - 32 IP - 4 DP - 2003 Apr TI - [What effect does the hydroxyapatite coating have in cementless knee arthroplasty?]. PG - 323-30 AB - The goal of this study was to compare the migration of noncemented diffusion sintered titanium fibermesh-coated tibial components with (HA group) and without (non-HA group) additional hydroxyapatite coating. For this purpose digital radiostereometry (DIRSA) was used to compare the migration after 2 and 9 years for the two groups. After 2 years the mean maximum subsidence of the HA-coated components (0.24+/-0.18 mm) was about one-half of the mean maximum subsidence of the non-HA-coated group (0.55+/-0.55 mm). After 9 years the maximum subsidence of the HA-coated components was still smaller, but not as pronounced as before. The same trend was also found for the endpoint maximum total point motion (MTPM). After 2 years the mean MTPM of the HA-coated components was 0.66+/-0.38 mm and of the non-HA group 0.73+/-0.50 mm. After 9 years the mean MTPM for the HA-coated components was 0.54+/-0.15 mm and for the non-HA-coated components 0.74+/-0.20 mm. None of the HA-coated tibial components but one of the non-HA group had to be revised and exchanged due to aseptic loosening. FAU - Hildebrand, R AU - Hildebrand R AD - Klinik fur Orthopadie und Rheumatologie, Philipps-Universitat, Marburg. hildebr2@post.med.uni-marburg.de FAU - Trappmann, D AU - Trappmann D FAU - Georg, C AU - Georg C FAU - Muller, H H AU - Muller HH FAU - Koller, M AU - Koller M FAU - Klose, K J AU - Klose KJ FAU - Griss, P AU - Griss P FAU - Kienapfel, H AU - Kienapfel H LA - ger PT - Clinical Trial PT - Comparative Study PT - English Abstract PT - Journal Article PT - Randomized Controlled Trial TT - Welchen Effekt hat die Hydroxylapatitbeschichtung beim zementfreien Kniegelenkersatz? Eine prospektiv randomisierte RSA-Langzeitstudie uber 9 Jahre. PL - Germany TA - Orthopade JT - Der Orthopade JID - 0331266 RN - 0 (Coated Materials, Biocompatible) RN - 91D9GV0Z28 (Durapatite) SB - IM MH - Aged MH - Aged, 80 and over MH - *Coated Materials, Biocompatible MH - *Durapatite MH - *Equipment Failure Analysis MH - Female MH - Follow-Up Studies MH - Humans MH - Image Processing, Computer-Assisted MH - *Knee Prosthesis MH - Male MH - Microscopy, Electron, Scanning MH - Middle Aged MH - Photogrammetry MH - Postoperative Complications/pathology/surgery MH - Prospective Studies MH - Prosthesis Design MH - Reoperation MH - Surface Properties EDAT- 2003/04/23 05:00 MHDA- 2003/11/13 05:00 CRDT- 2003/04/23 05:00 PHST- 2003/04/23 05:00 [pubmed] PHST- 2003/11/13 05:00 [medline] PHST- 2003/04/23 05:00 [entrez] AID - 10.1007/s00132-002-0443-5 [doi] PST - ppublish SO - Orthopade. 2003 Apr;32(4):323-30. doi: 10.1007/s00132-002-0443-5.