PMID- 16964615 OWN - NLM STAT- MEDLINE DCOM- 20070119 LR - 20150525 IS - 0740-3194 (Print) IS - 0740-3194 (Linking) VI - 56 IP - 4 DP - 2006 Oct TI - High-resolution continuously acquired peripheral MR angiography featuring partial parallel imaging GRAPPA. PG - 859-65 AB - Continuously-moving-table MRI, in contrast to traditional multistation techniques, potentially can improve the scan time efficiency of whole-body applications and provide seamless images of an extended field of view (FOV). Contrast-enhanced MR angiography (CE-MRA) in particular requires high spatial resolution and at the same time has rigid scan time constraints due to the limited arterial contrast window. In this study a reconstruction method for continuously acquired 3D data sets during table movement was combined with a self-calibrated partial parallel imaging algorithm (generalized autocalibrating partially parallel acquisitions (GRAPPA)). The method was applied to peripheral CE-MRA and compared with a standard continuously-moving-table MRA protocol. The gain in scan time was used to increase the data acquisition matrix and decrease the slice thickness. The method was evaluated in five healthy volunteers and applied to one patient with peripheral arterial occlusive disease (PAOD). The protocols were intraindividually compared with respect to the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in selected vessel segments, as well as overall vessel depiction. The combination of the continuously-moving-table technique with parallel imaging enabled the acquisition of seamless peripheral 3D MRA with increased resolution and an overall crisper appearance. FAU - Zenge, Michael O AU - Zenge MO AD - Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany. michael.zenge@uni-duisburg-essen.de FAU - Vogt, Florian M AU - Vogt FM FAU - Brauck, Katja AU - Brauck K FAU - Jokel, Michaela AU - Jokel M FAU - Barkhausen, Joerg AU - Barkhausen J FAU - Kannengiesser, Stephan AU - Kannengiesser S FAU - Ladd, Mark E AU - Ladd ME FAU - Quick, Harald H AU - Quick HH LA - eng PT - Journal Article PL - United States TA - Magn Reson Med JT - Magnetic resonance in medicine JID - 8505245 RN - 0 (Contrast Media) SB - IM MH - Algorithms MH - Contrast Media MH - Humans MH - Image Processing, Computer-Assisted MH - Imaging, Three-Dimensional MH - Leg/*blood supply MH - Magnetic Resonance Angiography/*methods MH - Peripheral Vascular Diseases/*diagnosis EDAT- 2006/09/12 09:00 MHDA- 2007/01/20 09:00 CRDT- 2006/09/12 09:00 PHST- 2006/09/12 09:00 [pubmed] PHST- 2007/01/20 09:00 [medline] PHST- 2006/09/12 09:00 [entrez] AID - 10.1002/mrm.21033 [doi] PST - ppublish SO - Magn Reson Med. 2006 Oct;56(4):859-65. doi: 10.1002/mrm.21033.