PMID- 17369268 OWN - NLM STAT- MEDLINE DCOM- 20070807 LR - 20070418 IS - 1099-5129 (Print) IS - 1099-5129 (Linking) VI - 9 IP - 5 DP - 2007 May TI - Unusual cause for an increase of the sensing integrity counter in a patient with inappropriate implantable cardioverter-defibrillator therapy. PG - 275-7 AB - We describe the case of a patient who presented with multiple implantable cardioverter-defibrillator (ICD) shock discharges 12 months after device implantation. Upon device interrogation, intermittent oversensing of electrical noise and potential ICD lead failure were suggested by a significant increase in the sensing integrity counter (SIC), a cumulative count of very short ventricular sensed intervals. Analysis of stored episodes, however, revealed that inappropriate ICD therapy had been caused by intermittent T-wave oversensing (TWO), and that the increase of the SIC resulted from the coincidence of TWO and premature ventricular complexes (PVCs). T-wave oversensing resolved and the SIC did not increase any more during follow-up after adjustment of ventricular sensitivity. The coincidence of TWO and PVCs should therefore be considered as an uncommon cause for short ventricular sensed intervals in ICD patients presenting with a suspect increase in the SIC. FAU - Vollmann, Dirk AU - Vollmann D AD - Abteilung Kardiologie und Pneumologie, Herzzentrum, Klinikum der Georg-August Universitat Gottingen, Robert-Koch-Strasse 40, 37075 Gottingen, Germany. dirkvollmann2000@aol.com FAU - Luthje, Lars AU - Luthje L FAU - Zabel, Markus AU - Zabel M LA - eng PT - Case Reports PT - Journal Article DEP - 20070316 PL - England TA - Europace JT - Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology JID - 100883649 SB - IM MH - Aged MH - Defibrillators, Implantable/*adverse effects MH - *Electrocardiography MH - Electromyography MH - Heart Ventricles/*physiopathology MH - Humans MH - Male EDAT- 2007/03/21 09:00 MHDA- 2007/08/08 09:00 CRDT- 2007/03/21 09:00 PHST- 2007/03/21 09:00 [pubmed] PHST- 2007/08/08 09:00 [medline] PHST- 2007/03/21 09:00 [entrez] AID - eum028 [pii] AID - 10.1093/europace/eum028 [doi] PST - ppublish SO - Europace. 2007 May;9(5):275-7. doi: 10.1093/europace/eum028. Epub 2007 Mar 16.