PMID- 20460950 OWN - NLM STAT- MEDLINE DCOM- 20101019 LR - 20100714 IS - 1423-0372 (Electronic) IS - 1011-6125 (Linking) VI - 88 IP - 4 DP - 2010 TI - Using Monte-Carlo-simulated radiation transport to calculate dose distribution in rats before irradiation with Leksell Gamma Knife 4C: technical note. PG - 208-15 LID - 10.1159/000314355 [doi] AB - BACKGROUND: Gamma knife surgery (GKS) is used at subnecrotic doses for temporal lobe epilepsy (TLE) treatment. Rat models of TLE have been used to probe the mechanisms underlying GKS. Previous GKS studies on rats have used the Leksell GammaPlan (LGP) treatment planning system to determine the irradiation time to achieve the dose to deliver. Since LGP is not designed for such small structures, it is important to calibrate the system for the rat brain. METHODS: We have used a Monte Carlo simulation (MCS) radiation transport scheme, with CT data as anatomical and tissue-specific information, to simulate the dose distribution in a rat brain when using a Leksell Gamma Knife. RESULTS: We show how dose distributions obtained by MCS quantitatively compare to those predicted by LGP, and discuss whether LGP should be used for studies involving rats. The energy deposited when using the 4-mm collimators was calculated for targets on both sides of the rat brain in the dorsal hippocampus, which allowed us to determine the exact time to irradiate rats with a given dose. CONCLUSION: The MCS method used in this study can easily be used for future GKS studies on small animals when accurate dose distributions are required. CI - 2010 S. Karger AG, Basel. FAU - Marcelin, Beatrice AU - Marcelin B AD - INSERM U751, Universite de la Mediterranee, Hopital La Timone, Marseille, France. FAU - Kjall, Per AU - Kjall P FAU - Johansson, Jonas AU - Johansson J FAU - Lundin, Anders AU - Lundin A FAU - Nordstrom, Hakan AU - Nordstrom H FAU - Eriksson, Markus AU - Eriksson M FAU - Bernard, Christophe AU - Bernard C FAU - Regis, Jean AU - Regis J LA - eng PT - Journal Article PT - Research Support, Non-U.S. Gov't DEP - 20100512 PL - Switzerland TA - Stereotact Funct Neurosurg JT - Stereotactic and functional neurosurgery JID - 8902881 SB - IM MH - Animals MH - Calibration MH - *Radiation Dosage MH - Radiosurgery/*instrumentation/methods MH - Radiotherapy Planning, Computer-Assisted/instrumentation/*methods MH - Rats EDAT- 2010/05/13 06:00 MHDA- 2010/10/20 06:00 CRDT- 2010/05/13 06:00 PHST- 2009/07/31 00:00 [received] PHST- 2010/03/04 00:00 [accepted] PHST- 2010/05/13 06:00 [entrez] PHST- 2010/05/13 06:00 [pubmed] PHST- 2010/10/20 06:00 [medline] AID - 000314355 [pii] AID - 10.1159/000314355 [doi] PST - ppublish SO - Stereotact Funct Neurosurg. 2010;88(4):208-15. doi: 10.1159/000314355. Epub 2010 May 12.