PMID- 22047294 OWN - NLM STAT- PubMed-not-MEDLINE DCOM- 20120223 LR - 20111103 IS - 1089-7623 (Electronic) IS - 0034-6748 (Linking) VI - 82 IP - 10 DP - 2011 Oct TI - Detection efficiency vs. cathode and anode separation in cylindrical vacuum photodiodes used for measuring x-rays from plasma focus device. PG - 103507 LID - 10.1063/1.3652962 [doi] AB - A qualitative study on the performance of cylindrical vacuum photodiodes (VPDs) for x-ray detection in plasma focus device has been carried out. Various parameters of VPD such as electrode's diameter, electrode's separation, and its sensitivity are experimentally tested in plasma focus environment. For the first time it is found experimentally that the electrode-separation in the lateral direction of the two coaxial electrodes of cylindrical VPD also plays an important role to increase the efficiency of the detector. The efficiency is found to be highest for the detector with smaller cathode-anode lateral gap (1.5 mm) with smaller photo cathode diameter (10 mm). A comparison between our VPD with PIN (BPX-65) diode as an x-ray detector has also been made. FAU - Borthakur, T K AU - Borthakur TK AD - Centre of Plasma Physics-Institute for Plasma Research, Tepesia, Sonapur, Assam, Pin-782 402, India. tkborthakur@yahoo.co.uk FAU - Talukdar, N AU - Talukdar N FAU - Neog, N K AU - Neog NK FAU - Rao, C V S AU - Rao CV FAU - Shyam, A AU - Shyam A LA - eng PT - Journal Article PL - United States TA - Rev Sci Instrum JT - The Review of scientific instruments JID - 0405571 EDAT- 2011/11/04 06:00 MHDA- 2011/11/04 06:01 CRDT- 2011/11/04 06:00 PHST- 2011/11/04 06:00 [entrez] PHST- 2011/11/04 06:00 [pubmed] PHST- 2011/11/04 06:01 [medline] AID - 10.1063/1.3652962 [doi] PST - ppublish SO - Rev Sci Instrum. 2011 Oct;82(10):103507. doi: 10.1063/1.3652962.