PMID- 22775367 OWN - NLM STAT- MEDLINE DCOM- 20130212 LR - 20151119 IS - 1557-8593 (Electronic) IS - 1520-9156 (Linking) VI - 14 IP - 10 DP - 2012 Oct TI - Evaluation of a new reusable insulin pen (ClikSTAR) in Canadian patients with type 1 and type 2 diabetes mellitus receiving insulin glargine. PG - 926-35 AB - BACKGROUND: The objective of this Canadian observational study was to assess the safety and patient satisfaction with the ClikSTAR((R)) (sanofi-aventis, Frankfurt am Main, Germany) reusable insulin pen in clinical practice. SUBJECTS AND METHODS: Patients with diabetes (n=2,517) were recruited from 103 sites, provided with the ClikSTAR pen, and instructed to report product technical complaints (PTCs), product technical failures (PTFs), and adverse events (AEs) over 6-12 weeks of use. A patient subgroup (n = 301) completed a pen use questionnaire at 12 weeks. The risk acceptance criterion was that no validated PTF led to a serious AE (SAE). PTFs were PTCs having a confirmed technical deficiency. RESULTS: Patients (68.5% with type 2 diabetes) had a mean age of 56.2 years, and 92.5% were insulin pen users. In total, 84 PTCs were reported by 79 (3.1%) patients. Most PTCs were due to pen handling errors. PTCs from 12 patients were possibly related to AEs, three of which were SAEs; none was confirmed to be due to a PTF. The single reported PTF was not related to an AE/SAE. In the substudy, 97.0% and 95.3% of patients rated "ease of learning" and "ease of use" of the ClikSTAR pen as excellent or good, respectively. Mean scores for patient satisfaction, convenience, flexibility, and recommendations of current treatment on the Diabetes Treatment Satisfaction Questionnaire (change version) had positive changes ranging from 1.0 to 1.8. CONCLUSIONS: With ClikSTAR, PTCs were infrequent, and there were no PTFs associated with AEs, demonstrating that the pen is reliable and safe with high levels of patient acceptance and satisfaction. FAU - Gottesman, Irving AU - Gottesman I AD - University of Toronto, Mississauga, Ontario, Canada. irvgottesman@hotmail.com FAU - Perron, Patrice AU - Perron P FAU - Berard, Lori AU - Berard L FAU - Stewart, John AU - Stewart J FAU - Basso, Nils AU - Basso N FAU - Mettimano, Kim AU - Mettimano K FAU - Elliott, Thomas AU - Elliott T LA - eng PT - Journal Article PT - Multicenter Study PT - Research Support, Non-U.S. Gov't DEP - 20120709 PL - United States TA - Diabetes Technol Ther JT - Diabetes technology & therapeutics JID - 100889084 RN - 0 (Hypoglycemic Agents) RN - 0 (Insulin, Long-Acting) RN - 2ZM8CX04RZ (Insulin Glargine) SB - IM MH - Canada/epidemiology MH - Diabetes Mellitus, Type 1/*drug therapy/epidemiology MH - Diabetes Mellitus, Type 2/*drug therapy/epidemiology MH - Drug Delivery Systems/*instrumentation MH - Equipment Design MH - Female MH - Humans MH - Hypoglycemic Agents/*administration & dosage MH - Injections, Subcutaneous/instrumentation MH - Insulin Glargine MH - Insulin, Long-Acting/*administration & dosage MH - Male MH - Middle Aged MH - Patient Satisfaction/*statistics & numerical data MH - Prospective Studies MH - Surveys and Questionnaires EDAT- 2012/07/11 06:00 MHDA- 2013/02/13 06:00 CRDT- 2012/07/11 06:00 PHST- 2012/07/11 06:00 [entrez] PHST- 2012/07/11 06:00 [pubmed] PHST- 2013/02/13 06:00 [medline] AID - 10.1089/dia.2012.0067 [doi] PST - ppublish SO - Diabetes Technol Ther. 2012 Oct;14(10):926-35. doi: 10.1089/dia.2012.0067. Epub 2012 Jul 9.