PMID- 22843372 OWN - NLM STAT- MEDLINE DCOM- 20130125 LR - 20220408 IS - 1349-9157 (Electronic) IS - 0449-3060 (Print) IS - 0449-3060 (Linking) VI - 53 IP - 5 DP - 2012 Sep TI - Quality assurance of MLC leaf position accuracy and relative dose effect at the MLC abutment region using an electronic portal imaging device. PG - 798-806 LID - 10.1093/jrr/rrs038 [doi] AB - We investigated an electronic portal image device (EPID)-based method to see whether it provides effective and accurate relative dose measurement at abutment leaves in terms of positional errors of the multi-leaf collimator (MLC) leaf position. A Siemens ONCOR machine was used. For the garden fence test, a rectangular field (0.2 20 cm) was sequentially irradiated 11 times at 2-cm intervals. Deviations from planned leaf positions were calculated. For the nongap test, relative doses at the MLC abutment region were evaluated by sequential irradiation of a rectangular field (2 20 cm) 10 times with a MLC separation of 2 cm without a leaf gap. The integral signal in a region of interest was set to position A (between leaves) and B (neighbor of A). A pixel value at position B was used as background and the pixel ratio (A/B 100) was calculated. Both tests were performed at four gantry angles (0, 90, 180 and 270 degrees ) four times over 1 month. For the nongap test the difference in pixel ratio between the first and last period was calculated. Regarding results, average deviations from planned positions with the garden fence test were within 0.5 mm at all gantry angles, and at gantry angles of 90 and 270 degrees tended to decrease gradually over the month. For the nongap test, pixel ratio tended to increase gradually in all leaves, leading to a decrease in relative doses at abutment regions. This phenomenon was affected by both gravity arising from the gantry angle, and the hardware-associated contraction of field size with this type of machine. FAU - Sumida, Iori AU - Sumida I AD - Department of Oral and Maxillofacial Radiology, Osaka University Graduate School of Dentistry, 1-8 Yamada-oka, Suita, Osaka, 565-0871 Japan. isumida@dent.osaka-u.ac.jp FAU - Yamaguchi, Hajime AU - Yamaguchi H FAU - Kizaki, Hisao AU - Kizaki H FAU - Koizumi, Masahiko AU - Koizumi M FAU - Ogata, Toshiyuki AU - Ogata T FAU - Takahashi, Yutaka AU - Takahashi Y FAU - Yoshioka, Yasuo AU - Yoshioka Y LA - eng PT - Journal Article DEP - 20120710 PL - England TA - J Radiat Res JT - Journal of radiation research JID - 0376611 SB - IM MH - Humans MH - Neoplasms/radiotherapy MH - Quality Assurance, Health Care/methods MH - Radiotherapy Planning, Computer-Assisted/instrumentation/standards/statistics & numerical data MH - Radiotherapy, Intensity-Modulated/instrumentation/*standards/statistics & numerical data PMC - PMC3430416 EDAT- 2012/07/31 06:00 MHDA- 2013/01/26 06:00 PMCR- 2012/07/10 CRDT- 2012/07/31 06:00 PHST- 2012/07/31 06:00 [entrez] PHST- 2012/07/31 06:00 [pubmed] PHST- 2013/01/26 06:00 [medline] PHST- 2012/07/10 00:00 [pmc-release] AID - rrs038 [pii] AID - 10.1093/jrr/rrs038 [doi] PST - ppublish SO - J Radiat Res. 2012 Sep;53(5):798-806. doi: 10.1093/jrr/rrs038. Epub 2012 Jul 10.