PMID- 23314254 OWN - NLM STAT- MEDLINE DCOM- 20140128 LR - 20211021 IS - 1998-3689 (Electronic) IS - 0301-4738 (Print) IS - 0301-4738 (Linking) VI - 61 IP - 5 DP - 2013 May TI - Segmentation error and macular thickness measurements obtained with spectral-domain optical coherence tomography devices in neovascular age-related macular degeneration. PG - 213-7 LID - 10.4103/0301-4738.97075 [doi] AB - PURPOSE: To evaluate frequency and severity of segmentation errors of two spectral-domain optical coherence tomography (SD-OCT) devices and error effect on central macular thickness (CMT) measurements. MATERIALS AND METHODS: Twenty-seven eyes of 25 patients with neovascular age-related macular degeneration, examined using the Cirrus HD-OCT and Spectralis HRA + OCT, were retrospectively reviewed. Macular cube 512 x 128 and 5-line raster scans were performed with the Cirrus and 512 x 25 volume scans with the Spectralis. Frequency and severity of segmentation errors were compared between scans. RESULTS: Segmentation error frequency was 47.4% (baseline), 40.7% (1 month), 40.7% (2 months), and 48.1% (6 months) for the Cirrus, and 59.3%, 62.2%, 57.8%, and 63.7%, respectively, for the Spectralis, differing significantly between devices at all examinations (P < 0.05), except at baseline. Average error score was 1.21 +/- 1.65 (baseline), 0.79 +/- 1.18 (1 month), 0.74 +/- 1.12 (2 months), and 0.96 +/- 1.11 (6 months) for the Cirrus, and 1.73 +/- 1.50, 1.54 +/- 1.35, 1.38 +/- 1.40, and 1.49 +/- 1.30, respectively, for the Spectralis, differing significantly at 1 month and 2 months (P < 0.02). Automated and manual CMT measurements by the Spectralis were larger than those by the Cirrus. CONCLUSIONS: The Cirrus HD-OCT had a lower frequency and severity of segmentation error than the Spectralis HRA + OCT. SD-OCT error should be considered when evaluating retinal thickness. FAU - Kim, Moosang AU - Kim M AD - Department of Ophthalmology, Kyung Hee University Hospital, Seoul, Korea. FAU - Lee, Seung Jun AU - Lee SJ FAU - Han, Jisang AU - Han J FAU - Yu, Seung-Young AU - Yu SY FAU - Kwak, Hyung Woo AU - Kwak HW LA - eng PT - Comparative Study PT - Journal Article PL - India TA - Indian J Ophthalmol JT - Indian journal of ophthalmology JID - 0405376 SB - IM MH - Aged MH - Female MH - Humans MH - Macula Lutea/*pathology MH - Macular Degeneration/complications/*diagnosis MH - Male MH - Retinal Neovascularization/*diagnosis/etiology MH - Retrospective Studies MH - Tomography, Optical Coherence/*methods PMC - PMC3730504 COIS- Conflict of Interest: None declared. EDAT- 2013/01/15 06:00 MHDA- 2014/01/29 06:00 PMCR- 2013/05/01 CRDT- 2013/01/15 06:00 PHST- 2013/01/15 06:00 [entrez] PHST- 2013/01/15 06:00 [pubmed] PHST- 2014/01/29 06:00 [medline] PHST- 2013/05/01 00:00 [pmc-release] AID - 97075 [pii] AID - IJO-61-213 [pii] AID - 10.4103/0301-4738.97075 [doi] PST - ppublish SO - Indian J Ophthalmol. 2013 May;61(5):213-7. doi: 10.4103/0301-4738.97075.