PMID- 24038218 OWN - NLM STAT- MEDLINE DCOM- 20140422 LR - 20211021 IS - 1098-2825 (Electronic) IS - 0887-8013 (Print) IS - 0887-8013 (Linking) VI - 27 IP - 5 DP - 2013 Sep TI - A simple, fast correction method of triglyceride interference in blood hemoglobin automated measurement. PG - 341-5 LID - 10.1002/jcla.21568 [doi] AB - BACKGROUND: To establish a reliable correction method for automated hemoglobin (HGB) measurement by minimizing the interference from blood high triglyceride (TG). METHODS: Fifty whole blood samples and 50 plasma samples containing variable TG concentrations were used to determine the centrifugation speed and time. Complete blood cell counts (CBCs) were performed by an automated hematology analyzer for 102 blood samples, in which high-level TG were artificially added. The same blood samples were centrifuged at low -speed to separate the plasma from blood cells. Then the plasma was analyzed by the same analyzer. By using the two CBC results, a correction formula was established to calculate the corrected HGB, mean corpuscular hemoglobin (MCH) and mean corpuscular hemoglobin concentration (MCHC) values. Comparisons were also made of HGB, MCH, and MCHC values before and after correction of in-patient individuals who received intralipid and developed lipemia. RESULTS: The percentage differences between the corrected and true values of HGB, MCH and MCHC were -0.28%, 0.06%, and -0.31%, respectively. The correlation coefficients of corrected values versus true values of HGB, MCH, and MCHC were 0.989, 0.935, and 0.717, respectively. This correction method was also effective for native lipemic samples. CONCLUSION: High blood TG level can cause blood turbidity and erroneously high HGB results by hematology analyzers commonly used in clinical laboratories. Adding a simple step of low-speed centrifugation and measurement of HGB in the plasma fraction allows a quick correction of HGB measurement in lipemic blood samples. CI - (c) 2012 Wiley Periodicals, Inc. FAU - Zeng, Su-Gen AU - Zeng SG AD - Department of Laboratory Medicine, West China Hospital, Sichuan University, Chengdu, P. R. China. FAU - Zeng, Ting-Ting AU - Zeng TT FAU - Jiang, Hong AU - Jiang H FAU - Wang, Lan-Lan AU - Wang LL FAU - Tang, Shu-Qiang AU - Tang SQ FAU - Sun, Yu-Ming AU - Sun YM FAU - Ying, Bin-Wu AU - Ying BW FAU - Jia, Yong-Qian AU - Jia YQ LA - eng PT - Journal Article PL - United States TA - J Clin Lab Anal JT - Journal of clinical laboratory analysis JID - 8801384 RN - 0 (Hemoglobins) RN - 0 (Triglycerides) SB - IM MH - *Automation, Laboratory MH - Blood Cell Count MH - Centrifugation MH - *Erythrocyte Indices MH - Hemoglobins/analysis MH - Humans MH - Hypertriglyceridemia/*blood MH - Triglycerides/*blood/chemistry PMC - PMC6807499 OTO - NOTNLM OT - correction method OT - hemoglobin concentration OT - high triglyceride EDAT- 2013/09/17 06:00 MHDA- 2014/04/23 06:00 PMCR- 2013/09/13 CRDT- 2013/09/17 06:00 PHST- 2012/07/29 00:00 [received] PHST- 2012/11/05 00:00 [revised] PHST- 2012/11/19 00:00 [accepted] PHST- 2013/09/17 06:00 [entrez] PHST- 2013/09/17 06:00 [pubmed] PHST- 2014/04/23 06:00 [medline] PHST- 2013/09/13 00:00 [pmc-release] AID - JCLA21568 [pii] AID - 10.1002/jcla.21568 [doi] PST - ppublish SO - J Clin Lab Anal. 2013 Sep;27(5):341-5. doi: 10.1002/jcla.21568.