PMID- 24532110 OWN - NLM STAT- MEDLINE DCOM- 20150105 LR - 20211021 IS - 1572-8595 (Electronic) IS - 1383-875X (Linking) VI - 39 IP - 3 DP - 2014 Apr TI - Incidence of nonphysiologic short VV intervals detected by the sensing integrity counter with integrated bipolar compared with true bipolar leads: clinically inconsequential or cause for concern? PG - 281-5 LID - 10.1007/s10840-013-9866-7 [doi] AB - INTRODUCTION: The need to detect impending implantable cardiac defibrillator (ICD) lead failure has grown. Automated sensing diagnostics have been developed for this reason. The sensing integrity counter (SIC) is one such oversensing diagnostic, which forms an integral part of the Medtronic lead integrity alert (LIA) feature on implantable defibrillators. It records nonphysiologic short VV intervals (NPSVVIs). It is unclear whether SIC data derived from integrated bipolar (IBP) leads need to be interpreted differently when compared to true bipolar (TBP) leads. We hypothesized that IBP ICD leads by virtue of a larger "antennae" may generate more NPSVVIs on than TBP leads, leading to more false-positive SIC counts. METHODS: Equal durations of remote monitoring records of 44 patients (mean age of 65.9 +/- 2.2 years, 52 % female) with IBP ICD leads and Medtronic (MDT) generators (IBP group) were compared with those of 44 randomly selected patients (64.0 +/- 2.2 years, 24 % female) who had TBP ICD leads and MDT generators (TBP group). Mean surveillance time, defined as the time over which the cumulative SIC count was acquired, was 614 +/- 44 days (TBP group) vs. 620 +/- 49 days (IBP group, p = ns). The mean time of follow-up following the first documented short VV interval was 115.2 months in the integrated bipolar group and 66.9 months in the true bipolar group. Leads on advisory were excluded from the study. RESULTS: A total of 26/44 patients in the IBP group displayed NPSVVI compared to 11/44 patients in the TBP group (59 vs. 25 %; p = 0.002, Fisher exact test). When adjusted for gender and lead age, the difference was still significant (p = 0.008). When evaluating the clinical consequence of NPSVVI in this cohort, 3/11 TBP leads with NPSVVI of >0 were eventually extracted due to additional abnormalities vs. 0/26 IBP leads with NPSVVI (p = 0.02, Fisher exact test). None of the IBP group patients with NPSVVI have developed inappropriate therapy from lead noise or a need for abandonment or extraction. CONCLUSION: Integrated bipolar ICD leads are more likely to have elevated SIC counts than true bipolar leads despite revealing no other evidence of lead failure. There does not appear to be a need for heightened surveillance in IBP leads with observed elevated SIC counts that have no other findings to suggest lead malfunction. FAU - Ng, Justin AU - Ng J AD - Brigham and Women's Hospital, Boston, MA, USA, jng123456@gmail.com. FAU - Chopra, Nagesh AU - Chopra N FAU - Barbhaiya, Chirag AU - Barbhaiya C FAU - Reichlin, Tobias AU - Reichlin T FAU - Nof, Eyal AU - Nof E FAU - Nagashima, Koichi AU - Nagashima K FAU - Tadros, Thomas AU - Tadros T FAU - Koplan, Bruce A AU - Koplan BA LA - eng PT - Comparative Study PT - Journal Article DEP - 20140216 PL - Netherlands TA - J Interv Card Electrophysiol JT - Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing JID - 9708966 SB - IM MH - Aged MH - *Defibrillators, Implantable MH - Equipment Design MH - Equipment Failure Analysis/*methods MH - Female MH - Humans MH - Male EDAT- 2014/02/18 06:00 MHDA- 2015/01/06 06:00 CRDT- 2014/02/18 06:00 PHST- 2013/08/21 00:00 [received] PHST- 2013/12/04 00:00 [accepted] PHST- 2014/02/18 06:00 [entrez] PHST- 2014/02/18 06:00 [pubmed] PHST- 2015/01/06 06:00 [medline] AID - 10.1007/s10840-013-9866-7 [doi] PST - ppublish SO - J Interv Card Electrophysiol. 2014 Apr;39(3):281-5. doi: 10.1007/s10840-013-9866-7. Epub 2014 Feb 16.