PMID- 25069780 OWN - NLM STAT- MEDLINE DCOM- 20150529 LR - 20181202 IS - 1496-8975 (Electronic) IS - 0832-610X (Linking) VI - 61 IP - 10 DP - 2014 Oct TI - Effects of fresh gas flow, tidal volume, and charcoal filters on the washout of sevoflurane from the Datex Ohmeda (GE) Aisys, Aestiva/5, and Excel 210 SE Anesthesia Workstations. PG - 935-42 LID - 10.1007/s12630-014-0200-3 [doi] AB - PURPOSE: We investigated the effects of tidal volume (VT), fresh gas flow (FGF), and a charcoal filter in the inspiratory limb on the washout of sevoflurane from the following Datex Ohmeda (GE) Anesthesia Workstations (AWSs): Aisys, Aestiva/5, and Excel 210SE. METHODS: After equilibrating the AWSs with 2% sevoflurane, the anesthetic was discontinued, and the absorbent anesthesia breathing circuit (ABC), reservoir bag, and test lung were changed. The lung was ventilated with 350 or 200 mL.breath(-1), 15 breaths.min(-1), and a FGF of 10 L.min(-1) while the washout of sevoflurane was performed in triplicate using a calibrated Datex Ohmeda Capnomac Ultima and a calibrated MIRAN SapphIRe XL ambient air analyzer until the concentration was /= 50 ppm. Placement of a charcoal filter in the inspiratory limb reduced the sevoflurane concentration to < 2 ppm in the Aisys and Aestiva/5 AWSs within two minutes. CONCLUSION: The GE AWSs should be purged with large FGFs and VTs ~350 mL.breath(-1) for ~25 min to achieve 10 ppm sevoflurane. The FGF should be maintained to avoid a rebound in anesthetic concentration. Charcoal filters rapidly decrease the anesthetic concentration to < 2 ppm. FAU - Sabouri, A Sassan AU - Sabouri AS AD - Department of Anesthesia, Women and Children's Hospital of Buffalo, SUNY at Buffalo, 219 Bryant ST, Buffalo, NY, 14222, USA. FAU - Lerman, Jerrold AU - Lerman J FAU - Heard, Christopher AU - Heard C LA - eng PT - Journal Article PT - Research Support, Non-U.S. Gov't DEP - 20140729 PL - United States TA - Can J Anaesth JT - Canadian journal of anaesthesia = Journal canadien d'anesthesie JID - 8701709 RN - 0 (Anesthetics, Inhalation) RN - 0 (Methyl Ethers) RN - 16291-96-6 (Charcoal) RN - 38LVP0K73A (Sevoflurane) SB - IM MH - Anesthesiology/*instrumentation MH - Anesthetics, Inhalation/*analysis MH - Charcoal/chemistry MH - Equipment Design MH - Filtration MH - Humans MH - Methyl Ethers/*analysis MH - Sevoflurane MH - Tidal Volume MH - Time Factors EDAT- 2014/07/30 06:00 MHDA- 2015/05/30 06:00 CRDT- 2014/07/30 06:00 PHST- 2014/01/30 00:00 [received] PHST- 2014/06/18 00:00 [accepted] PHST- 2014/07/30 06:00 [entrez] PHST- 2014/07/30 06:00 [pubmed] PHST- 2015/05/30 06:00 [medline] AID - 10.1007/s12630-014-0200-3 [doi] PST - ppublish SO - Can J Anaesth. 2014 Oct;61(10):935-42. doi: 10.1007/s12630-014-0200-3. Epub 2014 Jul 29.