PMID- 25281622 OWN - NLM STAT- MEDLINE DCOM- 20150616 LR - 20141004 IS - 0974-5181 (Electronic) IS - 0971-9784 (Linking) VI - 17 IP - 4 DP - 2014 Oct-Dec TI - Comparison between continuous non-invasive estimated cardiac output by pulse wave transit time and thermodilution method. PG - 273-7 LID - 10.4103/0971-9784.142059 [doi] AB - AIMS AND OBJECTIVES: Cardiac output (CO) measurement is essential for many therapeutic decisions in anesthesia and critical care. Most available non-invasive CO measuring methods have an invasive component. We investigate "pulse wave transit time" (estimated continuous cardiac output [esCCO]) a method of CO measurement that has no invasive component to its use. MATERIALS AND METHODS: After institutional ethical committee approval, 14 adult (21-85 years) patients undergoing surgery and requiring pulmonary artery catheter (PAC) for measuring CO, were included. Postoperatively CO readings were taken simultaneously with thermodilution (TD) via PAC and esCCO, whenever a change in CO was expected due to therapeutic interventions. Both monitoring methods were continued until patients' discharge from the Intensive Care Unit and observer recording values using TD method was blinded to values measured by esCCO system. RESULTS: Three hundred and one readings were obtained simultaneously from both methods. Correlation and concordance between the two methods was derived using Bland-Altman analysis. Measured values showed significant correlation between esCCO and TD ( r = 0.6, P < 0.001, 95% confidence limits of 0.51-0.68). Mean and (standard deviation) for bias and precision were 0.13 (2.27) L/min and 6.56 (2.19) L/min, respectively. The 95% confidence interval for bias was - 4.32 to 4.58 L/min and for precision 2.27 to10.85 L/min. CONCLUSIONS: Although, esCCO is the only true non-invasive continuous CO monitor available and even though its values change proportionately to TD method (gold standard) with the present degree of error its utility for clinical/therapeutic decision-making is questionable. FAU - Sinha, Ashish C AU - Sinha AC AD - Department of Anaesthesia, Drexel University College of Medicine, Philadelphia, PA, USA. FAU - Singh, Preet Mohinder AU - Singh PM FAU - Grewal, Navneet AU - Grewal N FAU - Aman, Mansoor AU - Aman M FAU - Dubowitz, Gerald AU - Dubowitz G LA - eng PT - Comparative Study PT - Journal Article PL - India TA - Ann Card Anaesth JT - Annals of cardiac anaesthesia JID - 9815987 SB - IM CIN - Ann Card Anaesth. 2014 Oct-Dec;17(4):278. PMID: 25281623 MH - Adult MH - Aged MH - Aged, 80 and over MH - Cardiac Output/*physiology MH - Catheterization, Swan-Ganz/methods MH - Female MH - Humans MH - Male MH - Middle Aged MH - Monitoring, Intraoperative/methods/statistics & numerical data MH - Monitoring, Physiologic/*methods/*statistics & numerical data MH - Prospective Studies MH - Pulse Wave Analysis/*methods/*statistics & numerical data MH - Reproducibility of Results MH - Thermodilution/methods/statistics & numerical data MH - Young Adult EDAT- 2014/10/05 06:00 MHDA- 2015/06/17 06:00 CRDT- 2014/10/05 06:00 PHST- 2014/10/05 06:00 [entrez] PHST- 2014/10/05 06:00 [pubmed] PHST- 2015/06/17 06:00 [medline] AID - AnnCardAnaesth_2014_17_4_273_142059 [pii] AID - 10.4103/0971-9784.142059 [doi] PST - ppublish SO - Ann Card Anaesth. 2014 Oct-Dec;17(4):273-7. doi: 10.4103/0971-9784.142059.