PMID- 26305422 OWN - NLM STAT- MEDLINE DCOM- 20160726 LR - 20191210 IS - 1502-7686 (Electronic) IS - 0036-5513 (Linking) VI - 75 IP - 7 DP - 2015 Nov TI - Hypo- and hypernatremia results in inaccurate erythrocyte mean corpuscular volume measurement in vitro, when using Sysmex XE 2100. PG - 588-94 LID - 10.3109/00365513.2015.1062534 [doi] AB - INTRODUCTION: Automated hematology analyzers dilute patient erythrocytes with an isoosmotic diluent before quantitating the erythrocyte mean cell volume (MCV). However, if patient plasma osmolality differs from the diluent, water will cross the erythrocytes membrane and establish a new equilibrium across the membrane. Since the new equilibrium is reached before the measurement of the MCV, the measured MCV may not reflect the true MCV in vivo. AIM: Calculation of the theoretical change in MCV at changed P-Sodium/P-Osmolality and to investigate if the automated blood cell counter Sysmex XE 2100 measures MCV correctly in hypo- and hyperosmolality and hypo-and hypernatremia. In addition, to examine whether the theoretically calculated change in MCV corresponds with the experimentally determined MCV change. METHOD: Theoretical calculation of the MCV inaccuracy at hypo- and hypernatremia, as well as at hypo- and hyperosmolality. Experimental studies with comparison of MCV measured at Sysmex XE 2100 to MCV found by using the manual measured packed cell volume method. RESULTS AND CONCLUSION: Measurement of MCV in hypo- and hypernatremia patients using the automated blood cell counter Sysmex XE 2100 resulted in inaccurate MCV. The experimental results also revealed a strong correlation between P-Osmolality/P-Sodium and MCV inaccuracy (R(2) = 0.70/0.85) similar to the theoretically calculated MCV inaccuracy. We suggest using mean cellular Hb (MCH) instead of MCV, mean corpuscular Hb concentration (MCHC) and B-Erythrocyte volume fraction (EVF). Alternatively, we suggest standardizing the measured MCV to a normal P-Sodium e.g. 140 mmol/L to estimate the in vivo MCV. FAU - Philipsen, Jens Peter AU - Philipsen JP AD - a Nordsjaellands Hospital, Biomedical Laboratory , Hillerod , Denmark. FAU - Madsen, Kirsten Vikkelso AU - Madsen KV LA - eng PT - Evaluation Study PT - Journal Article DEP - 20150814 PL - England TA - Scand J Clin Lab Invest JT - Scandinavian journal of clinical and laboratory investigation JID - 0404375 SB - IM MH - *Erythrocyte Indices MH - Female MH - Hematologic Tests/*standards MH - Humans MH - Hypernatremia/blood/*complications MH - Hyponatremia/blood/*complications MH - Male MH - Reproducibility of Results OTO - NOTNLM OT - Anemias OT - cell size OT - erythrocytes OT - hematology OT - hyponatremia EDAT- 2015/08/26 06:00 MHDA- 2016/07/28 06:00 CRDT- 2015/08/26 06:00 PHST- 2015/08/26 06:00 [entrez] PHST- 2015/08/26 06:00 [pubmed] PHST- 2016/07/28 06:00 [medline] AID - 10.3109/00365513.2015.1062534 [doi] PST - ppublish SO - Scand J Clin Lab Invest. 2015 Nov;75(7):588-94. doi: 10.3109/00365513.2015.1062534. Epub 2015 Aug 14.