PMID- 26328983 OWN - NLM STAT- MEDLINE DCOM- 20160601 LR - 20170214 IS - 2473-4209 (Electronic) IS - 0094-2405 (Linking) VI - 42 IP - 9 DP - 2015 Sep TI - Fully iterative scatter corrected digital breast tomosynthesis using GPU-based fast Monte Carlo simulation and composition ratio update. PG - 5342-55 LID - 10.1118/1.4928139 [doi] AB - PURPOSE: In digital breast tomosynthesis (DBT), scatter correction is highly desirable, as it improves image quality at low doses. Because the DBT detector panel is typically stationary during the source rotation, antiscatter grids are not generally compatible with DBT; thus, a software-based scatter correction is required. This work proposes a fully iterative scatter correction method that uses a novel fast Monte Carlo simulation (MCS) with a tissue-composition ratio estimation technique for DBT imaging. METHODS: To apply MCS to scatter estimation, the material composition in each voxel should be known. To overcome the lack of prior accurate knowledge of tissue composition for DBT, a tissue-composition ratio is estimated based on the observation that the breast tissues are principally composed of adipose and glandular tissues. Using this approximation, the composition ratio can be estimated from the reconstructed attenuation coefficients, and the scatter distribution can then be estimated by MCS using the composition ratio. The scatter estimation and image reconstruction procedures can be performed iteratively until an acceptable accuracy is achieved. For practical use, (i) the authors have implemented a fast MCS using a graphics processing unit (GPU), (ii) the MCS is simplified to transport only x-rays in the energy range of 10-50 keV, modeling Rayleigh and Compton scattering and the photoelectric effect using the tissue-composition ratio of adipose and glandular tissues, and (iii) downsampling is used because the scatter distribution varies rather smoothly. RESULTS: The authors have demonstrated that the proposed method can accurately estimate the scatter distribution, and that the contrast-to-noise ratio of the final reconstructed image is significantly improved. The authors validated the performance of the MCS by changing the tissue thickness, composition ratio, and x-ray energy. The authors confirmed that the tissue-composition ratio estimation was quite accurate under a variety of conditions. Our GPU-based fast MCS implementation took approximately 3 s to generate each angular projection for a 6 cm thick breast, which is believed to make this process acceptable for clinical applications. In addition, the clinical preferences of three radiologists were evaluated; the preference for the proposed method compared to the preference for the convolution-based method was statistically meaningful (p < 0.05, McNemar test). CONCLUSIONS: The proposed fully iterative scatter correction method and the GPU-based fast MCS using tissue-composition ratio estimation successfully improved the image quality within a reasonable computational time, which may potentially increase the clinical utility of DBT. FAU - Kim, Kyungsang AU - Kim K AD - Bio Imaging and Signal Processing Laboratory, Department of Bio and Brain Engineering, KAIST 291, Daehak-ro, Yuseong-gu, Daejeon 34141, South Korea. FAU - Lee, Taewon AU - Lee T AD - Medical Imaging and Radiotherapeutics Laboratory, Department of Nuclear and Quantum Engineering, KAIST 291, Daehak-ro, Yuseong-gu, Daejeon 34141, South Korea. FAU - Seong, Younghun AU - Seong Y AD - Samsung Advanced Institute of Technology, Samsung Electronics, 130, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 443-803, South Korea. FAU - Lee, Jongha AU - Lee J AD - Samsung Advanced Institute of Technology, Samsung Electronics, 130, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 443-803, South Korea. FAU - Jang, Kwang Eun AU - Jang KE AD - Samsung Advanced Institute of Technology, Samsung Electronics, 130, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 443-803, South Korea. FAU - Choi, Jaegu AU - Choi J AD - Korea Electrotechnology Research Institute (KERI), 111, Hanggaul-ro, Sangnok-gu, Ansan-si, Gyeonggi-do, 426-170, South Korea. FAU - Choi, Young Wook AU - Choi YW AD - Korea Electrotechnology Research Institute (KERI), 111, Hanggaul-ro, Sangnok-gu, Ansan-si, Gyeonggi-do, 426-170, South Korea. FAU - Kim, Hak Hee AU - Kim HH AD - Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine, 88 Olympic-ro, 43-gil, Songpa-gu, Seoul, 138-736, South Korea. FAU - Shin, Hee Jung AU - Shin HJ AD - Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine, 88 Olympic-ro, 43-gil, Songpa-gu, Seoul, 138-736, South Korea. FAU - Cha, Joo Hee AU - Cha JH AD - Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine, 88 Olympic-ro, 43-gil, Songpa-gu, Seoul, 138-736, South Korea. FAU - Cho, Seungryong AU - Cho S AD - Medical Imaging and Radiotherapeutics Laboratory, Department of Nuclear and Quantum Engineering, KAIST 291, Daehak-ro, Yuseong-gu, Daejeon 34141, South Korea. FAU - Ye, Jong Chul AU - Ye JC AD - Bio Imaging and Signal Processing Laboratory, Department of Bio and Brain Engineering, KAIST 291, Daehak-ro, Yuseong-gu, Daejeon 34141, South Korea. LA - eng PT - Journal Article PT - Research Support, Non-U.S. Gov't PL - United States TA - Med Phys JT - Medical physics JID - 0425746 SB - IM MH - *Computer Graphics MH - Humans MH - Image Processing, Computer-Assisted/*methods MH - Mammography/*methods MH - *Monte Carlo Method MH - Radiographic Image Enhancement/*methods MH - *Scattering, Radiation MH - Time Factors EDAT- 2015/09/04 06:00 MHDA- 2016/06/02 06:00 CRDT- 2015/09/03 06:00 PHST- 2015/09/03 06:00 [entrez] PHST- 2015/09/04 06:00 [pubmed] PHST- 2016/06/02 06:00 [medline] AID - 10.1118/1.4928139 [doi] PST - ppublish SO - Med Phys. 2015 Sep;42(9):5342-55. doi: 10.1118/1.4928139.