PMID- 26365343 OWN - NLM STAT- MEDLINE DCOM- 20160512 LR - 20151012 IS - 1364-5528 (Electronic) IS - 0003-2654 (Linking) VI - 140 IP - 21 DP - 2015 Nov 7 TI - An 8 minute colorimetric paper-based reverse phase vertical flow serum microarray for screening of hyper IgE syndrome. PG - 7327-34 LID - 10.1039/c5an01013f [doi] AB - Reverse phase microarrays are useful tools for affinity-based detection in hundreds of samples simultaneously. However, current methods typically require long assay times and fluorescent detection. Here we describe a paper-based Vertical Flow Microarray (VFM) assay as a rapid 8-minute colorimetric alternative for reverse phase microarray analysis. The VFM platform was optimized for detection of IgE with a detection limit of 1.9 mug mL(-1) in whole serum. Optimized conditions were then used to screen 113 serum samples simultaneously for hyper IgE syndrome (hIgE), a rare primary immunodeficiency characterized by elevated levels of IgE. The same set of samples were then analysed with a conventional planar microarray with fluorescent detection for head-to-head testing. Both assays found elevated levels in three out of four hIgE patient samples, whereas no control samples displayed elevated levels in either method. The comparison experiments showed a good correlation between the two assays, as determined from a linear correlation study (Pearson's r = 0.76). Further, the assay-time reduction and reproducibility (intra assay CV = 12.4 +/- 4.11%) demonstrate the applicability of the VFM platform for high throughput reverse phase screening. FAU - Reutersward, Philippa AU - Reutersward P AD - Division of Proteomics and Nanobiotechnology, Science for Life Laboratory, KTH Royal Institute of Technology, Sweden. helene.andersson.svahn@scilifelab.se. FAU - Gantelius, Jesper AU - Gantelius J FAU - Andersson Svahn, Helene AU - Andersson Svahn H LA - eng PT - Journal Article PT - Research Support, Non-U.S. Gov't PL - England TA - Analyst JT - The Analyst JID - 0372652 RN - 37341-29-0 (Immunoglobulin E) RN - 9004-70-0 (Collodion) SB - IM MH - Area Under Curve MH - Collodion/chemistry MH - Colorimetry/*instrumentation/*methods MH - Equipment Design MH - Humans MH - Immunoglobulin E/*analysis MH - Job Syndrome/*blood/*diagnosis MH - Limit of Detection MH - Microarray Analysis MH - Paper MH - Protein Array Analysis/methods MH - Reference Values MH - Reproducibility of Results MH - Sensitivity and Specificity EDAT- 2015/09/15 06:00 MHDA- 2016/05/14 06:00 CRDT- 2015/09/15 06:00 PHST- 2015/09/15 06:00 [entrez] PHST- 2015/09/15 06:00 [pubmed] PHST- 2016/05/14 06:00 [medline] AID - 10.1039/c5an01013f [doi] PST - ppublish SO - Analyst. 2015 Nov 7;140(21):7327-34. doi: 10.1039/c5an01013f.