PMID- 26547433 OWN - NLM STAT- MEDLINE DCOM- 20160912 LR - 20151207 IS - 1873-4235 (Electronic) IS - 0956-5663 (Linking) VI - 77 DP - 2016 Mar 15 TI - Functionalized gold nanoclusters as fluorescent labels for immunoassays: Application to human serum immunoglobulin E determination. PG - 1055-61 LID - S0956-5663(15)30336-5 [pii] LID - 10.1016/j.bios.2015.08.011 [doi] AB - A quantitative immunoassay for the determination of immunoglobulin E (IgE) in human serum using gold nanoclusters (AuNCs) as fluorescent label was developed. Water soluble AuNCs were synthesized using lipoic acid and then thoroughly characterized. The obtained AuNCs have a particle size of 2.7 +/- 0.1 nm and maximum fluorescence emission at 710 nm. The synthesized AuNCs showed very good stability of the fluorescent signal with light exposure and at neutral and slightly basic media. A covalent bioconjugation of these AuNCs with the desired antibody was carried out by the carbodiimide reaction. After due optimization of such bioconjugation reaction, a molar ratio 1:3 (antibody:AuNCs) was selected. The bioconjugate maintained an intense luminescence emission, slightly red-shifted as compared to the free AuNCs. Two typical immunoassay configurations, competitive and sandwich, were assayed and their performance for IgE determination critically compared. After the different immunoassay steps were accomplished, the fluorescence emission of the bioconjugate was measured. While the sandwich format provided a detection limit (DL) of 10 ng/mL and a linear range between 25 and 565 ng/mL of IgE, the competitive format revealed a DL of 0.2 ng/mL with a linear range between 0.3 and 7.1 ng/mL The applicability of the more sensitive competitive fluorescent immunoassay was assessed by successful analysis of the IgE in human serum and comparison of results with those from a commercial kit. The main advantages of the proposed AuNCs-based fluorimetric method include a low DL and a simple immunoassay protocol involving few reagents. CI - Copyright (c) 2015. Published by Elsevier B.V. FAU - Alonso, Maria Cruz AU - Alonso MC AD - Department of Physical and Analytical Chemistry, Faculty of Chemistry, University of Oviedo, Julian Claveria, 8, 33006 Oviedo, Spain. FAU - Trapiella-Alfonso, Laura AU - Trapiella-Alfonso L AD - Department of Physical and Analytical Chemistry, Faculty of Chemistry, University of Oviedo, Julian Claveria, 8, 33006 Oviedo, Spain. FAU - Fernandez, Jose M Costa AU - Fernandez JM AD - Department of Physical and Analytical Chemistry, Faculty of Chemistry, University of Oviedo, Julian Claveria, 8, 33006 Oviedo, Spain. Electronic address: jcostafe@uniovi.es. FAU - Pereiro, Rosario AU - Pereiro R AD - Department of Physical and Analytical Chemistry, Faculty of Chemistry, University of Oviedo, Julian Claveria, 8, 33006 Oviedo, Spain. Electronic address: mrpereiro@uniovi.es. FAU - Sanz-Medel, Alfredo AU - Sanz-Medel A AD - Department of Physical and Analytical Chemistry, Faculty of Chemistry, University of Oviedo, Julian Claveria, 8, 33006 Oviedo, Spain. LA - eng PT - Journal Article PT - Research Support, Non-U.S. Gov't DEP - 20150809 PL - England TA - Biosens Bioelectron JT - Biosensors & bioelectronics JID - 9001289 RN - 0 (Antibodies, Monoclonal) RN - 0 (Fluorescent Dyes) RN - 0 (Nanoconjugates) RN - 37341-29-0 (Immunoglobulin E) RN - 7440-57-5 (Gold) SB - IM MH - Antibodies, Monoclonal/chemistry MH - Biosensing Techniques/*instrumentation MH - Blood Chemical Analysis/*instrumentation MH - Equipment Design MH - Equipment Failure Analysis MH - Fluorescent Dyes/chemical synthesis MH - Gold/*chemistry MH - Humans MH - Immunoassay/*instrumentation MH - Immunoglobulin E/*blood MH - Metal Nanoparticles/*chemistry/ultrastructure MH - Nanoconjugates/chemistry/ultrastructure MH - Reproducibility of Results MH - Sensitivity and Specificity OTO - NOTNLM OT - Competitive immunoassay OT - Fluorescence OT - Gold nanoclusters OT - Human serum OT - Immunoglobulin E EDAT- 2015/11/09 06:00 MHDA- 2016/09/13 06:00 CRDT- 2015/11/09 06:00 PHST- 2015/06/03 00:00 [received] PHST- 2015/07/30 00:00 [revised] PHST- 2015/08/08 00:00 [accepted] PHST- 2015/11/09 06:00 [entrez] PHST- 2015/11/09 06:00 [pubmed] PHST- 2016/09/13 06:00 [medline] AID - S0956-5663(15)30336-5 [pii] AID - 10.1016/j.bios.2015.08.011 [doi] PST - ppublish SO - Biosens Bioelectron. 2016 Mar 15;77:1055-61. doi: 10.1016/j.bios.2015.08.011. Epub 2015 Aug 9.