PMID- 26612266 OWN - NLM STAT- MEDLINE DCOM- 20161103 LR - 20191210 IS - 1944-0057 (Electronic) IS - 1944-0057 (Linking) VI - 33 IP - 3 DP - 2016 TI - Evaluation and validation of a multi-residue method based on biochip technology for the simultaneous screening of six families of antibiotics in muscle and aquaculture products. PG - 403-19 LID - 10.1080/19440049.2015.1125529 [doi] AB - The Evidence Investigator system (Randox, UK) is a biochip and semi-automated system. The microarray kit II (AM II) is capable of detecting several compounds belonging to different families of antibiotics: quinolones, ceftiofur, thiamphenicol, streptomycin, tylosin and tetracyclines. The performance of this innovative system was evaluated for the detection of antibiotic residues in new matrices, in muscle of different animal species and in aquaculture products. The method was validated according to the European Decision No. EC/2002/657 and the European guideline for the validation of screening methods, which represents a complete initial validation. The false-positive rate was equal to 0% in muscle and in aquaculture products. The detection capabilities CCbeta for 12 validated antibiotics (enrofloxacin, difloxacin, ceftiofur, desfuroyl ceftiofur cysteine disulfide, thiamphenicol, florfenicol, tylosin, tilmicosin, streptomycin, dihydrostreptomycin, tetracycline, doxycycline) were all lower than the respective maximum residue limits (MRLs) in muscle from different animal origins (bovine, ovine, porcine, poultry). No cross-reactions were observed with other antibiotics, neither with the six detected families nor with other families of antibiotics. The AM II kit could be applied to aquaculture products but with higher detection capabilities from those in muscle. The detection capabilities CCbeta in aquaculture products were respectively at 0.25, 0.10 and 0.5 of the respective MRL in aquaculture products for enrofloxacin, tylosin and oxytetracycline. The performance of the AM II kit has been compared with other screening methods and with the performance characteristics previously determined in honey. FAU - Gaudin, Valerie AU - Gaudin V AD - a European Union Reference Laboratory, Anses Fougeres , La Haute Marche - BP 90203 , F-35302 Fougeres Cedex , France. FAU - Hedou, Celine AU - Hedou C AD - a European Union Reference Laboratory, Anses Fougeres , La Haute Marche - BP 90203 , F-35302 Fougeres Cedex , France. FAU - Soumet, Christophe AU - Soumet C AD - a European Union Reference Laboratory, Anses Fougeres , La Haute Marche - BP 90203 , F-35302 Fougeres Cedex , France. FAU - Verdon, Eric AU - Verdon E AD - a European Union Reference Laboratory, Anses Fougeres , La Haute Marche - BP 90203 , F-35302 Fougeres Cedex , France. LA - eng PT - Evaluation Study PT - Journal Article PT - Validation Study DEP - 20160125 PL - England TA - Food Addit Contam Part A Chem Anal Control Expo Risk Assess JT - Food additives & contaminants. Part A, Chemistry, analysis, control, exposure & risk assessment JID - 101485040 RN - 0 (Anti-Bacterial Agents) SB - IM MH - Animals MH - Anti-Bacterial Agents/*analysis/*classification MH - *Aquaculture MH - Automation MH - Cattle MH - Drug Residues/*analysis MH - Food Contamination/*analysis MH - Honey/analysis MH - *Microarray Analysis MH - Muscles/*chemistry MH - Salmon MH - Seafood/*analysis MH - Sheep MH - Swine OTO - NOTNLM OT - European Union Decision No. 2002/657/EC OT - Validation OT - antibiotic OT - aquaculture OT - biochip OT - multi-array OT - muscle OT - screening EDAT- 2015/11/28 06:00 MHDA- 2016/11/04 06:00 CRDT- 2015/11/28 06:00 PHST- 2015/11/28 06:00 [entrez] PHST- 2015/11/28 06:00 [pubmed] PHST- 2016/11/04 06:00 [medline] AID - 10.1080/19440049.2015.1125529 [doi] PST - ppublish SO - Food Addit Contam Part A Chem Anal Control Expo Risk Assess. 2016;33(3):403-19. doi: 10.1080/19440049.2015.1125529. Epub 2016 Jan 25.