PMID- 27210737 OWN - NLM STAT- MEDLINE DCOM- 20170517 LR - 20170517 IS - 1873-3476 (Electronic) IS - 0378-5173 (Linking) VI - 514 IP - 2 DP - 2016 Dec 5 TI - Exploring the impact of sample flowrate on in vitro measurements of metered dose inhaler performance. PG - 420-427 LID - S0378-5173(16)30405-7 [pii] LID - 10.1016/j.ijpharm.2016.05.025 [doi] AB - Pharmacopoeial methods for measurement of the aerodynamic particle size distribution (APSD) of metered dose inhalers (MDIs) by cascade impaction specify a sampling flow rate of 28.3L/min. However, there is little data within the literature to rationalize this figure, or to support its clinical relevance. In addition, the standard United States Pharmacopoeia Induction Port (USP IP) used for testing is known to inaccurately reflect deposition behavior in the upper airway, further compromising the relevance of testing, for product development. This article describes experimental studies of the effect of sampling flow rate on APSD data gathered using an Andersen Cascade Impactor (ACI). Tests were carried out using two different formulations to assess the influence of formulation composition. In addition, comparative testing with an Alberta Idealised Throat, in place of the USP IP, to ensure more realistic representation of the upper airway. The results show how measured APSD and fine particle dose, the dose than on the basis of size would be expected to deposit in the lung, vary as a function of test methodology, providing insight as to how the testing can be modified towards greater clinical relevance. CI - Copyright (c) 2016 The Author(s). Published by Elsevier B.V. All rights reserved. FAU - Lewis, D A AU - Lewis DA AD - Chiesi Ltd., Bath Rd. Ind. Est., Chippenham, Wiltshire SN14 0AB, UK. Electronic address: d.lewis@chiesi.com. FAU - O'Shea, H AU - O'Shea H AD - Chiesi Ltd., Bath Rd. Ind. Est., Chippenham, Wiltshire SN14 0AB, UK. FAU - Church, T K AU - Church TK AD - Chiesi Ltd., Bath Rd. Ind. Est., Chippenham, Wiltshire SN14 0AB, UK. FAU - Brambilla, G AU - Brambilla G AD - Chiesi Ltd., Bath Rd. Ind. Est., Chippenham, Wiltshire SN14 0AB, UK. FAU - Traini, D AU - Traini D AD - Respiratory Technology, Woolcock Institute of Medical Research, The University of Sydney, 2006, Australia; Discipline of Pharmacology, Sydney Medical School, The University of Sydney, 2006, Australia. FAU - Young, P M AU - Young PM AD - Respiratory Technology, Woolcock Institute of Medical Research, The University of Sydney, 2006, Australia; Discipline of Pharmacology, Sydney Medical School, The University of Sydney, 2006, Australia. LA - eng PT - Journal Article DEP - 20160519 PL - Netherlands TA - Int J Pharm JT - International journal of pharmaceutics JID - 7804127 SB - IM MH - Administration, Inhalation MH - Drug Compounding MH - *Equipment Design MH - Humans MH - Metered Dose Inhalers/*standards MH - Particle Size MH - Rheology/*instrumentation OTO - NOTNLM OT - Aerodynamic particle size distribution OT - Alberta idealized throat OT - Cascade impactor OT - Fine particle dose OT - Metered dose inhalers OT - Testing conditions EDAT- 2016/05/24 06:00 MHDA- 2017/05/18 06:00 CRDT- 2016/05/24 06:00 PHST- 2016/03/23 00:00 [received] PHST- 2016/05/11 00:00 [revised] PHST- 2016/05/13 00:00 [accepted] PHST- 2016/05/24 06:00 [pubmed] PHST- 2017/05/18 06:00 [medline] PHST- 2016/05/24 06:00 [entrez] AID - S0378-5173(16)30405-7 [pii] AID - 10.1016/j.ijpharm.2016.05.025 [doi] PST - ppublish SO - Int J Pharm. 2016 Dec 5;514(2):420-427. doi: 10.1016/j.ijpharm.2016.05.025. Epub 2016 May 19.