PMID- 27455500 OWN - NLM STAT- MEDLINE DCOM- 20170504 LR - 20231112 IS - 1526-9914 (Electronic) IS - 1526-9914 (Linking) VI - 17 IP - 4 DP - 2016 Jul 8 TI - Electrocardiogram-gated coronary CT angiography dose estimates using ImPACT. PG - 342-356 LID - 10.1120/jacmp.v17i4.6218 [doi] AB - The primary study objective was to assess radiation doses using a modified form of the Imaging Performance Assessment of Computed Tomography (CT) scanner (ImPACT) patient dosimetry for cardiac applications on an Aquilion ONE ViSION Edition scanner, including the Ca score, target computed tomography angiography (CTA), prospective CTA, continuous CTA/cardiac function analysis (CFA), and CTA/CFA modulation. Accordingly, we clarified the CT dose index (CTDI) to determine the relationship between heart rate (HR) and X-ray exposure. As a secondary objective, we compared radiation doses using modified ImPACT, a whole-body dosimetry phantom study, and the k-factor method to verify the validity of the dose results obtained with modified ImPACT. The effective dose determined for the reference person (4.66 mSv at 60 beats per minute (bpm) and 33.43 mSv at 90bpm) were approximately 10% less than those determined for the phantom study (5.28 mSv and 36.68 mSv). The effective doses according to the k-factor (0.014 mSv*mGy-1*cm-1; 2.57 mSv and 17.10 mSv) were significantly lower than those obtained with the other two methods. In the present study, we have shown that ImPACT, when modified for cardiac applications, can assess both absorbed and effective doses. The results of our dose comparison indicate that modified ImPACT dose assessment is a promising and practical method for evaluating coronary CTA. CI - (c) 2016 The Authors FAU - Kobayashi, Masanao AU - Kobayashi M AD - Graduate School of Health Sciences, Fujita Health University, Toyoake, Japan. masa1121@fujita-hu.ac.jp. FAU - Asada, Yasuki AU - Asada Y FAU - Matsubara, Kosuke AU - Matsubara K FAU - Suzuki, Shouichi AU - Suzuki S FAU - Koshida, Kichiro AU - Koshida K FAU - Matsunaga, Yuta AU - Matsunaga Y FAU - Haba, Tomonobu AU - Haba T FAU - Kawaguchi, Ai AU - Kawaguchi A FAU - Toyama, Hiroshi AU - Toyama H FAU - Kato, Ryouichi AU - Kato R LA - eng PT - Journal Article DEP - 20160708 PL - United States TA - J Appl Clin Med Phys JT - Journal of applied clinical medical physics JID - 101089176 SB - IM MH - Cardiac-Gated Imaging Techniques/*methods MH - Coronary Angiography/*instrumentation MH - Electrocardiography/*instrumentation MH - Heart/*diagnostic imaging MH - Humans MH - *Phantoms, Imaging MH - Prospective Studies MH - Retrospective Studies MH - *Tomography Scanners, X-Ray Computed MH - Tomography, X-Ray Computed/*instrumentation PMC - PMC5690042 EDAT- 2016/07/28 06:00 MHDA- 2017/05/05 06:00 PMCR- 2016/07/08 CRDT- 2016/07/26 06:00 PHST- 2015/11/11 00:00 [received] PHST- 2016/02/23 00:00 [accepted] PHST- 2016/02/23 00:00 [revised] PHST- 2016/07/26 06:00 [entrez] PHST- 2016/07/28 06:00 [pubmed] PHST- 2017/05/05 06:00 [medline] PHST- 2016/07/08 00:00 [pmc-release] AID - ACM20342 [pii] AID - 10.1120/jacmp.v17i4.6218 [doi] PST - epublish SO - J Appl Clin Med Phys. 2016 Jul 8;17(4):342-356. doi: 10.1120/jacmp.v17i4.6218.