PMID- 28431009 OWN - NLM STAT- MEDLINE DCOM- 20190208 LR - 20190215 IS - 1532-2092 (Electronic) IS - 1099-5129 (Linking) VI - 20 IP - 5 DP - 2018 May 1 TI - Artefact-free late gadolinium enhancement imaging in patients with implanted cardiac devices using a modified broadband sequence: current strategies and results from a real-world patient cohort. PG - 801-807 LID - 10.1093/europace/eux016 [doi] AB - AIMS: Cardiac magnetic resonance (CMR) imaging in patients with implanted cardiac devices is often limited by device-related imaging artefacts. Limitations can potentially be overcome by employing a broadband late gadolinium enhancement (LGE)-CMR imaging technique. The purpose of this study was to investigate the relationship between implanted cardiac devices and the optimal frequency offset on broadband LGE-CMR imaging to increase the artefact-free visibility of myocardial segments. METHODS AND RESULTS: A phantom study was performed to characterize magnetic field disturbances related to 15 different cardiac devices. This was complemented by B0 and B1+ imaging of three different device types in four healthy volunteers. Findings were validated in 28 patients with an indication for arrhythmogenic substrate characterization before catheter ablation. In the phantom study, the placement of a PM, implantable cardioverter-defibrillator (ICD) or CRT-D generator led to a significant impairment of the radiofrequency field. B0 mapping in phantom and volunteers showed the highest off-resonance maximum with CRT-D systems with the maximum off-resonance significantly decreasing for ICD or PM systems, respectively. In all patients, with conventional LGE imaging 73.1% (61.5-92.3%) of LV segments were free of device-related artefacts, while with the broadband LGE technique, a significant increase of artefact-free segments was achieved [96.4% (85.7-100%); P = 0.00008]. CONCLUSION: Using a modified broadband sequence for LGE imaging significantly increased the number of artefact-free myocardial segments thereby leading to improved diagnostic value of the CMR exam. Since the occurrence and extent of hyperintensity artefacts are closely related to the individual device, more studies are warranted to evaluate if the results can be extrapolated to other devices and manufacturers. FAU - Hilbert, Sebastian AU - Hilbert S AD - Department of Electrophysiology, HELIOS Heart Center Leipzig, Leipzig, Germany. FAU - Weber, Alexander AU - Weber A AD - Department of Electrophysiology, HELIOS Heart Center Leipzig, Leipzig, Germany. FAU - Nehrke, Kay AU - Nehrke K AD - Philips Research Laboratories, Hamburg, Germany. FAU - Bornert, Peter AU - Bornert P AD - Philips Research Laboratories, Hamburg, Germany. FAU - Schnackenburg, Bernhard AU - Schnackenburg B AD - Philips Research Laboratories, Hamburg, Germany. FAU - Oebel, Sabrina AU - Oebel S AD - Department of Electrophysiology, HELIOS Heart Center Leipzig, Leipzig, Germany. FAU - Spampinato, Ricardo AU - Spampinato R AD - Department of Cardiac Surgery, HELIOS Heart Center Leipzig, Leipzig, Germany. FAU - Rogge, Cathleen AU - Rogge C AD - Department of Electrophysiology, HELIOS Heart Center Leipzig, Leipzig, Germany. FAU - Richter, Sergio AU - Richter S AD - Department of Electrophysiology, HELIOS Heart Center Leipzig, Leipzig, Germany. FAU - Hindricks, Gerhard AU - Hindricks G AD - Department of Electrophysiology, HELIOS Heart Center Leipzig, Leipzig, Germany. FAU - Paetsch, Ingo AU - Paetsch I AD - Department of Electrophysiology, HELIOS Heart Center Leipzig, Leipzig, Germany. FAU - Jahnke, Cosima AU - Jahnke C AD - Department of Electrophysiology, HELIOS Heart Center Leipzig, Leipzig, Germany. LA - eng PT - Journal Article PL - England TA - Europace JT - Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology JID - 100883649 RN - 0 (Contrast Media) RN - AU0V1LM3JT (Gadolinium) SB - IM MH - Adult MH - Aged MH - Arrhythmias, Cardiac/etiology/*prevention & control MH - Artifacts MH - Contrast Media/therapeutic use MH - *Defibrillators, Implantable MH - Female MH - Gadolinium/*therapeutic use MH - Humans MH - *Image Enhancement/instrumentation/methods MH - Kaplan-Meier Estimate MH - *Magnetic Resonance Imaging/instrumentation/methods MH - Male MH - Middle Aged MH - Phantoms, Imaging MH - Reproducibility of Results EDAT- 2017/04/22 06:00 MHDA- 2019/02/09 06:00 CRDT- 2017/04/22 06:00 PHST- 2016/08/07 00:00 [received] PHST- 2017/01/17 00:00 [accepted] PHST- 2017/04/22 06:00 [pubmed] PHST- 2019/02/09 06:00 [medline] PHST- 2017/04/22 06:00 [entrez] AID - 3739747 [pii] AID - 10.1093/europace/eux016 [doi] PST - ppublish SO - Europace. 2018 May 1;20(5):801-807. doi: 10.1093/europace/eux016.