PMID- 29024795 OWN - NLM STAT- MEDLINE DCOM- 20190114 LR - 20190114 IS - 1522-9629 (Electronic) IS - 1094-5539 (Linking) VI - 48 DP - 2018 Feb TI - Are valved holding chambers (VHCs) interchangeable? An in vitro evaluation of VHC equivalence. PG - 179-184 LID - S1094-5539(17)30184-0 [pii] LID - 10.1016/j.pupt.2017.10.005 [doi] AB - INTRODUCTION: The European Medicines Agency (EMA) requires that a specific valved holding chamber (VHC) is designated for use with a given pressurised metered dose inhaler (pMDI). No other regulatory authorities impose similar requirements, implying that VHCs are interchangeable. This in vitro study, employing EMA assessment criteria, assessed the equivalence of four anti-static VHCs (aVHCs) versus the non-conducting VHC most widely referenced in pMDI monographs, the AeroChamber Plus (AC+) VHC. MATERIAL & METHODS: The "reference" AC + VHC was prepared by soaking in detergent solution. The four test aVHCs (AeroChamber Plus Flow-Vu [AC + FV]; Compact Space Chamber Plus [CSC+]; InspiraChamber [IC]; OptiChamber Diamond [OCD]) were tested "out-of-packet". Twenty devices of each type were evaluated. A salbutamol pMDI was actuated into each VHC with a 2-s delay between actuation and Andersen Cascade Impactor (ACI) sampling. Drug deposition in four ACI particle size groups was assessed: Group 1, >5.8-10 mum; Group 2, >3.3-5.8 mum; Group 3, >1.1-3.3 mum; Group 4,