PMID- 29683797 OWN - NLM STAT- MEDLINE DCOM- 20190517 LR - 20210217 IS - 1437-4331 (Electronic) IS - 1434-6621 (Linking) VI - 56 IP - 9 DP - 2018 Aug 28 TI - Uncertainty evaluation in clinical chemistry, immunoassay, hematology and coagulation analytes using only external quality assessment data. PG - 1447-1457 LID - 10.1515/cclm-2017-1199 [doi] AB - BACKGROUND: Measurement uncertainty (MU) is a parameter associated with the result of a measurement that characterizes its dispersion. We report results for estimating MU following the application of a top-down procedure using only proficiency test data to establish uncertainty levels for various analytes. METHODS: Data were obtained from 142 laboratories participating in the Beijing Center for Clinical Laboratory (BCCL) proficiency testing/external quality assessment (PT/EQA) schemes. The 24-month study included six selected PT shipments to obtain estimates for 50th percentile (median) and 90th percentile MUs and to compare those estimates to usual analytic goals. The number of laboratory participants varied for each trial. The expanded uncertainty (U) was calculated using a cover factor of k=2 for a confidence interval of 95%. All reproducibility, method and laboratory biases came from the PT/EQA data. RESULTS: The median U (k=2) ranged from 3.2% (plasma sodium, indirect ion selective electrode) to 32.8% (triglycerides, free glycerol blanking) for clinical chemistry analyte means from participants in the same method group. Immunoassay analyte median U results ranged from 11.3% (CA125 tumor marker, Roche) to 33.8% (prostate-specific antigen [PSA], Abbott). The range for median U was 3.5% (red blood cell [RBC], Abx) to 30.3% (fibrinogen [FBG], other) for hematology and coagulation analytes. The MUs for most analytes satisfied quality requirements. CONCLUSIONS: The use of PT/EQA data, when available, provides an effective means for estimating uncertainties associated with quantitative measurements. Thus, medical laboratories can calculate their own MUs. Proficiency testing organizers can provide participants with an additional MU estimate using only EQA data, which may be updated at the end of each survey. FAU - Qin, Yanyan AU - Qin Y AD - Department of Clinical Laboratory, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, P.R. China. FAU - Zhou, Rui AU - Zhou R AD - Department of Clinical Laboratory, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, P.R. China. FAU - Wang, Wei AU - Wang W AD - Department of Blood Transfusion, Beijing Ditan Hospital, Capital Medical University, Beijing, P.R. China. FAU - Yin, Hongyi AU - Yin H AD - Department of Clinical Laboratory, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, P.R. China. FAU - Yang, Yanmin AU - Yang Y AD - Department of Clinical Laboratory, Fengtai Teaching Hospital of Capital Medical University, Beijing, P.R. China. FAU - Yue, Yuhong AU - Yue Y AD - Beijing Center for Clinical Laboratories, Beijing, P.R. China. FAU - Tong, Qing AU - Tong Q AD - Beijing Center for Clinical Laboratories, Beijing, P.R. China. FAU - Liu, Lu AU - Liu L AD - Beijing Center for Clinical Laboratories, Beijing, P.R. China. FAU - Jin, Yali AU - Jin Y AD - Beijing Center for Clinical Laboratories, Beijing, P.R. China. FAU - Shi, Yankun AU - Shi Y AD - Beijing Center for Clinical Laboratories, Beijing, P.R. China. FAU - Zhang, Shunli AU - Zhang S AD - Department of Clinical Laboratory, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, P.R. China. FAU - Zhang, Jianping AU - Zhang J AD - Department of Clinical Laboratory, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, P.R. China. FAU - Zhang, Rui AU - Zhang R AD - Department of Clinical Laboratory, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, P.R. China. FAU - Zuo, Chang AU - Zuo C AD - Department of Clinical Laboratory, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, P.R. China. FAU - Jia, Tingting AU - Jia T AD - Department of Clinical Laboratory, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, P.R. China. FAU - Wang, Ning AU - Wang N AD - Department of Clinical Laboratory, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, P.R. China. FAU - Wang, Qingtao AU - Wang Q AD - Department of Clinical Laboratory, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, P.R. China. AD - Beijing Center for Clinical Laboratories, No.8 Gongti South Street, Chaoyang District, Beijing, 100020, P.R. China, E-mail: wqt36@163.com. LA - eng PT - Journal Article PL - Germany TA - Clin Chem Lab Med JT - Clinical chemistry and laboratory medicine JID - 9806306 SB - IM MH - Blood Coagulation Tests/*standards MH - Clinical Chemistry Tests/*standards MH - Immunoassay/*standards MH - Laboratories/*standards MH - Quality Assurance, Health Care MH - Quality Control MH - Reproducibility of Results MH - *Uncertainty OTO - NOTNLM OT - bias OT - measurement uncertainty OT - proficiency testing/external quality assessment OT - reproducibility OT - standardized EDAT- 2018/04/24 06:00 MHDA- 2019/05/18 06:00 CRDT- 2018/04/24 06:00 PHST- 2017/12/22 00:00 [received] PHST- 2018/03/13 00:00 [accepted] PHST- 2018/04/24 06:00 [pubmed] PHST- 2019/05/18 06:00 [medline] PHST- 2018/04/24 06:00 [entrez] AID - cclm-2017-1199 [pii] AID - 10.1515/cclm-2017-1199 [doi] PST - ppublish SO - Clin Chem Lab Med. 2018 Aug 28;56(9):1447-1457. doi: 10.1515/cclm-2017-1199.