PMID- 30961573 OWN - NLM STAT- MEDLINE DCOM- 20190808 LR - 20200225 IS - 1471-2466 (Electronic) IS - 1471-2466 (Linking) VI - 19 IP - 1 DP - 2019 Apr 8 TI - Comparison of bench test results measuring the accuracy of peak flow meters. PG - 74 LID - 10.1186/s12890-019-0837-3 [doi] LID - 74 AB - BACKGROUND: The study evaluates and compares the accuracy of nine peak flow meters ("PFMs") and spirometers that are currently available in Europe and have Conformite Europeene ("CE") marking. The CE marking is a manufacturer's declaration that their product complies with European health regulations and it is a requirement for marketing medical devices in Europe. METHODS: The nine devices were selected as they all had received or were in the process of receiving CE approval in Europe and were readily obtainable. The devices were bench tested following the ISO 23747:2015 accuracy guidelines for medical devices measuring peak flow. All standards, including accuracy, from these guidelines must be met to obtain CE marking. This study was performed with a certified piston pump testing apparatus. The apparatus chosen was the pulmonary waveform generator manufactured by Piston Medical Ltd. Using predefined flow (time) and volume (time) waveforms, peak flow meters and spirometers were tested for validation and calibration. Three CE guideline tests were utilised, and standards require that all three tests are passed for the device to obtain certification. RESULTS: Of the nine devices that were tested, two passed and seven failed. The devices that passed the tests were the Smart Peak Flow(R) and the Mini Wright(R). CONCLUSIONS: A high percentage of devices failed accuracy testing in this study. This is a concern as the CE marking is a manufacturer's certification documenting the accuracy, reliability and safety of devices. Of the seven devices that failed all have the CE marking. All tested devices are on the market in Europe based upon studies conducted by each of the manufacturers. The data used to obtain CE certification of these devices, however, are not in the public domain. FAU - VanZeller, Cristiano AU - VanZeller C AD - Royal London Hospital, London, UK. FAU - Williams, Andrew AU - Williams A AD - Guy's Hospital, London, UK. FAU - Pollock, Ian AU - Pollock I AUID- ORCID: 0000-0001-6964-3424 AD - St Mary's Hospital, London, UK. ianpollock@doctors.org.uk. LA - eng GR - N/A/The study was funded by Smart Respiratory Products Limited/ PT - Comparative Study PT - Journal Article DEP - 20190408 PL - England TA - BMC Pulm Med JT - BMC pulmonary medicine JID - 100968563 MH - Europe MH - European Union MH - Forced Expiratory Volume MH - Humans MH - Lung/*physiology MH - Monitoring, Ambulatory/*instrumentation MH - Peak Expiratory Flow Rate MH - Reproducibility of Results MH - Spirometry/*instrumentation/standards PMC - PMC6454778 OTO - NOTNLM OT - Accuracy OT - Asthma OT - CE OT - Peak flow meter COIS- ETHICS APPROVAL AND CONSENT TO PARTICIPATE: Not applicable. CONSENT FOR PUBLICATION: Not applicable. COMPETING INTERESTS: The authors declare that they have no competing interests. PUBLISHER'S NOTE: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations. EDAT- 2019/04/10 06:00 MHDA- 2019/08/09 06:00 PMCR- 2019/04/08 CRDT- 2019/04/10 06:00 PHST- 2018/09/16 00:00 [received] PHST- 2019/03/24 00:00 [accepted] PHST- 2019/04/10 06:00 [entrez] PHST- 2019/04/10 06:00 [pubmed] PHST- 2019/08/09 06:00 [medline] PHST- 2019/04/08 00:00 [pmc-release] AID - 10.1186/s12890-019-0837-3 [pii] AID - 837 [pii] AID - 10.1186/s12890-019-0837-3 [doi] PST - epublish SO - BMC Pulm Med. 2019 Apr 8;19(1):74. doi: 10.1186/s12890-019-0837-3.