PMID- 31278545 OWN - NLM STAT- MEDLINE DCOM- 20210809 LR - 20210809 IS - 1573-2614 (Electronic) IS - 1387-1307 (Print) IS - 1387-1307 (Linking) VI - 34 IP - 3 DP - 2020 Jun TI - Validity and reliability of pocket-sized ultrasound devices in measurement of optic nerve sheath diameter in ICU patients. PG - 597-605 LID - 10.1007/s10877-019-00351-7 [doi] AB - The measurement of the optic nerve sheath diameter (ONSD) by ultrasonography (USG) is particularly important for intracranial pressure (ICP) monitoring when invasive measurements are not possible or are contraindicated. Standard USG (SUDs) devices can be bulky and may break down the workflow. The validity and reliability of pocket-sized USG devices (PSUDs) compared to SUDs for ONSD measurement has not been investigated, yet. We compared the reliability and validity of PSUDs with SUDs for ONSD measurement. 35 patients were included in the study. ONSD measurements with PSUDs and SUDs were performed by two clinicians as three different measurements in both sagittal and transverse axis of the eye. There was agreement between mean transverse ONSD (mtONSD), mean sagittal ONSD (msONSD), and mean ONSD(mONSD) values measured with each device according to Bland-Altman test (p = 0.166, p = 0.135, p = 0.066, respectively) with no proportional bias (p = 0.544, p = 0.868, p = 0.929, respectively). Intraclass correlation coefficients (ICC) were found to be greater than 0.7 for mtONSD, msONSD, and mONSD values measured by SUD (ICC = 0.790, ICC = 0.817, and ICC = 0.844, respectively). Similarly, for mtONSD, msONSD, and mONSD values measured by PSUD, ICC were found greater than 0.7 (ICC = 0.763, ICC = 0.814, and ICC = 0.843, respectively). There was no statistically significant difference between mtONSD and msONSD (p = 0.441 for SUD and p = 0.893 for PSUD). There was a good correlation between mtONSD and msONSD (r = 0.767 for SUD and r = 0.816 for PSUD). The total variance between three different measurements in the transverse and sagittal axis was +/- 0.6 mm. PSUDs can be used with similar validity and reliability as SUDs for ONSD measurement. FAU - Ozdemir, Ugur AU - Ozdemir U AUID- ORCID: 0000-0002-8738-3512 AD - Division of Critical Care Medicine, Department of Internal Medicine, Gazi University School of Medicine, Ankara, Turkey. ugureuzdemir@gmail.com. FAU - Cimen, Meltem AU - Cimen M AD - Division of Critical Care Medicine, Department of Anesthesiology, Gazi University School of Medicine, Ankara, Turkey. FAU - Guney, Tuba AU - Guney T AD - Division of Critical Care Medicine, Department of Neurology, Gazi University School of Medicine, Ankara, Turkey. FAU - Gursel, Gul AU - Gursel G AD - Department of Pulmonary Critical Care Medicine, Gazi University School of Medicine, Ankara, Turkey. LA - eng PT - Journal Article PT - Observational Study DEP - 20190705 PL - Netherlands TA - J Clin Monit Comput JT - Journal of clinical monitoring and computing JID - 9806357 SB - IM MH - Aged MH - Central Nervous System MH - Critical Care/*methods MH - Female MH - Humans MH - Intensive Care Units MH - Intracranial Hypertension MH - Intracranial Pressure/physiology MH - Male MH - Middle Aged MH - Monitoring, Intraoperative/*instrumentation/*methods MH - Monitoring, Physiologic MH - Optic Nerve/physiopathology/*surgery MH - Prospective Studies MH - ROC Curve MH - Reproducibility of Results MH - Ultrasonography/*instrumentation/*methods PMC - PMC7223003 OTO - NOTNLM OT - Intensive care unit OT - Intracranial pressure OT - Noninvasive monitoring OT - Optic nerve sheath OT - Pocket-sized ultrasonography device OT - Respiratory failure COIS- The authors declare that they have no conflicts of interest. EDAT- 2019/07/07 06:00 MHDA- 2021/08/10 06:00 PMCR- 2020/05/14 CRDT- 2019/07/07 06:00 PHST- 2019/02/12 00:00 [received] PHST- 2019/07/01 00:00 [accepted] PHST- 2019/07/07 06:00 [pubmed] PHST- 2021/08/10 06:00 [medline] PHST- 2019/07/07 06:00 [entrez] PHST- 2020/05/14 00:00 [pmc-release] AID - 10.1007/s10877-019-00351-7 [pii] AID - 351 [pii] AID - 10.1007/s10877-019-00351-7 [doi] PST - ppublish SO - J Clin Monit Comput. 2020 Jun;34(3):597-605. doi: 10.1007/s10877-019-00351-7. Epub 2019 Jul 5.