PMID- 31760240 OWN - NLM STAT- MEDLINE DCOM- 20200325 LR - 20200325 IS - 1879-1336 (Electronic) IS - 1054-8807 (Linking) VI - 44 DP - 2020 Jan-Feb TI - In vivo tissue reaction within the outflow conduit in patients supported by HeartWare HVAD. PG - 107156 LID - S1054-8807(19)30321-7 [pii] LID - 10.1016/j.carpath.2019.107156 [doi] AB - AIM: The frequency, extent, and nature of tissue ingrowth within the continuous-flow left ventricular assist device (cf-LVAD) outflow conduit has not been systematically assessed. We sought to characterize conduit histopathology at explantation in a cohort of patients with HeartWare ventricular assist device (HVAD) and assess the effect on pump performance. METHODS: Patients undergoing routine histopathological assessment of a HeartWare HVAD removed at transplantation or autopsy were assessed. Outflow conduits were examined macroscopically, and visible tissue was sectioned for microscopic evaluation. In patients who had undergone prior contrast-enhanced computerized tomography (CT) with HVAD in situ, the outflow conduit was measured at the aortic anastomosis and 5 cm proximal to the anastomosis, in the axial and sagittal planes. All patients had their pump flow, flow pulsatility, current, and speed determined from log files examined at 1, 3, 6, 9, and 12 months after LVAD implantation. RESULTS: Twenty-five consecutive patients were assessed (24 LVAD, 1 biventricular assist device (BiVAD)). Of the 26 outflow grafts assessed, there was evidence of tissue ingrowth reaction in 24 (92%) grafts. The most common site was the distal anastomosis (18/24, 75%), with the graft body involved in 14 of 24 (58%) grafts. Microscopic evaluation revealed acute inflammatory infiltrate in 4 of 24 grafts (17%), chronic inflammatory infiltrate in 14 of 24 (58%), neointima formation in 18 of 24 (75%) and fibrosis in 18 of 24 (75%) grafts. The median depth of tissue was 1 mm (range, 0-2 mm). The mean conduit diameter was 9.5 +/- 0.6 mm at the aortic anastomosis compared with 11.1 +/- 0.5 mm 5 cm proximal to the anastomosis (p < 0.0001). In patients with unchanged pump speed one month after implantation, analysis of log files revealed a significant (5.8 +/- 8.6%) decrease in pump flow (4.65 +/- 0.86 vs 4.38 +/- 0.92 L/min, p = 0.01) and flow pulsatility (5.00 +/- 1.10 vs 4.16 +/- 1.05 L/min, p = 0.006). CONCLUSIONS: There is evidence of tissue formation within the HVAD outflow conduit in the vast majority of patients, most commonly located at the aortic anastomosis. This is associated with significantly decreased pump flow over time. CI - Crown Copyright (c) 2019. Published by Elsevier Inc. All rights reserved. FAU - Jain, P AU - Jain P AD - Cardiology Department, St Vincent's Hospital, Sydney, Australia; University of New South Wales, Sydney, Australia. FAU - Robson, D AU - Robson D AD - Cardiology Department, St Vincent's Hospital, Sydney, Australia. FAU - Shehab, S AU - Shehab S AD - Victor Chang Cardiac Research Institute, Sydney, Australia. FAU - Muthiah, K AU - Muthiah K AD - Cardiology Department, St Vincent's Hospital, Sydney, Australia; Victor Chang Cardiac Research Institute, Sydney, Australia. FAU - Jansz, P AU - Jansz P AD - Cardiology Department, St Vincent's Hospital, Sydney, Australia. FAU - Qiu, M R AU - Qiu MR AD - Anatomical Pathology Department, St Vincent's Hospital, Sydney, Australia. FAU - Barrett, W AU - Barrett W AD - Anatomical Pathology Department, St Vincent's Hospital, Sydney, Australia. FAU - Sivasubramaniam, V AU - Sivasubramaniam V AD - Anatomical Pathology Department, St Vincent's Hospital, Sydney, Australia. FAU - Kumaradevan, N AU - Kumaradevan N AD - Anatomical Pathology Department, St Vincent's Hospital, Sydney, Australia. FAU - Macdonald, P S AU - Macdonald PS AD - Cardiology Department, St Vincent's Hospital, Sydney, Australia; University of New South Wales, Sydney, Australia; Victor Chang Cardiac Research Institute, Sydney, Australia. FAU - Hayward, C S AU - Hayward CS AD - Cardiology Department, St Vincent's Hospital, Sydney, Australia; University of New South Wales, Sydney, Australia; Victor Chang Cardiac Research Institute, Sydney, Australia. Electronic address: cshayward@stvincents.com.au. LA - eng PT - Journal Article DEP - 20191010 PL - United States TA - Cardiovasc Pathol JT - Cardiovascular pathology : the official journal of the Society for Cardiovascular Pathology JID - 9212060 SB - IM MH - Autopsy MH - Device Removal MH - Female MH - Foreign-Body Reaction/*etiology/pathology MH - Heart Failure/pathology/physiopathology/*therapy MH - *Heart-Assist Devices MH - Humans MH - Male MH - Middle Aged MH - Myocardium/*pathology MH - Prosthesis Design MH - Prosthesis Implantation/*adverse effects/*instrumentation MH - Retrospective Studies MH - Treatment Outcome MH - Ventricular Function, Left OTO - NOTNLM OT - Continuous-flow left ventricular assist device OT - Foreign body OT - Outflow conduit OT - Pump flow OT - Tissue reaction EDAT- 2019/11/25 06:00 MHDA- 2020/03/26 06:00 CRDT- 2019/11/25 06:00 PHST- 2019/08/01 00:00 [received] PHST- 2019/09/23 00:00 [revised] PHST- 2019/09/24 00:00 [accepted] PHST- 2019/11/25 06:00 [pubmed] PHST- 2020/03/26 06:00 [medline] PHST- 2019/11/25 06:00 [entrez] AID - S1054-8807(19)30321-7 [pii] AID - 10.1016/j.carpath.2019.107156 [doi] PST - ppublish SO - Cardiovasc Pathol. 2020 Jan-Feb;44:107156. doi: 10.1016/j.carpath.2019.107156. Epub 2019 Oct 10.