PMID- 32215725 OWN - NLM STAT- MEDLINE DCOM- 20210413 LR - 20210807 IS - 1435-702X (Electronic) IS - 0721-832X (Linking) VI - 258 IP - 7 DP - 2020 Jul TI - Reproducibility and agreement of four anterior segment-optical coherence tomography devices for anterior chamber angle measurements. PG - 1475-1481 LID - 10.1007/s00417-020-04648-1 [doi] AB - PURPOSE: To compare the reproducibility and agreement of anterior chamber angle (ACA) parameters and metrics obtained by four different anterior segment-optical coherence tomography (AS-OCT) devices. METHODS: In this prospective study, 30 eyes from 15 normal subjects underwent anterior segment angle scanning using the Spectralis, Cirrus, and Optovue spectral domain optical coherence tomography (SD-OCT), as well as the Visante time-domain optical coherence tomography (TD-OCT). For each eye, the scan line was performed perpendicularly on the inferior (270 degrees ) angle, and the inferior ACA image was acquired 2 times. Inter-instrument and intra-instrument, as well as inter-observer and intra-observer reproducibility of anterior chamber angle metrics, Schwalbe's line (SL) to scleral spur (SS) distance (TM-Span), angle opening distance (AOD), and trabecular iris space area (TISA) measurements, were evaluated by intraclass correlation coefficients (ICCs) and Bland-Altman plots with limits of agreement (LoA). RESULTS: For this cohort of 30 eyes of 15 normal subjects, the mean TM-Span, AOD, and TISA were 0.966 +/- 0.198 mm, 0.750 +/- 0.205 mm, and 0.286 +/- 0.090 mm(2) from the Spectralis; 0.929 +/- 0.113 mm, 0.717 +/- 0.120 mm, and 0.267 +/- 0.095 mm(2) from the Cirrus; 0.923 +/- 0.191 mm, 0.683 +/- 0.161 mm, and 0.265 +/- 0.072 mm(2) from the Optovue; and 0.970 +/- 0.070 mm, 0.705 +/- 0.150 mm, and 0.279 +/- 0.065 mm(2) from the Visante. The intra-instrument (ICCs > 0.838), intra-grader (ICCs > 0.910), and inter-grader (ICCs > 0.869) agreement were good. Agreement between the four instruments was also good with ICCs from 0.901 to 0.967 for TM-Span, 0.887 to 0.941 for AOD, and 0.923 to 0.961 for TISA. CONCLUSIONS: Consistent and reproducible ACA measurements could be obtained from multiple AS-OCT devices including both SD- and TD-OCT instruments. These findings have relevance when considering multiple imaging devices in future studies. FAU - Pan, Xiaojing AU - Pan X AD - Qingdao Eye Hospital, Shandong Eye Institute, Shandong First Medical University & Shandong Academy of Medical Sciences, Qingdao, Shandong, People's Republic of China. FAU - Maram, Jyotsna AU - Maram J AD - Doheny Image Reading Center, Doheny Eye Institute, University of California Los Angeles, Los Angeles, CA, USA. FAU - Nittala, Muneeswar G AU - Nittala MG AD - Doheny Image Reading Center, Doheny Eye Institute, University of California Los Angeles, Los Angeles, CA, USA. FAU - Francis, Brian A AU - Francis BA AD - Doheny Image Reading Center, Doheny Eye Institute; Department of Ophthalmology, David Geffen School of Medicine, University of California Los Angeles, Los Angeles, CA, USA. FAU - Chopra, Vikas AU - Chopra V AD - Doheny Image Reading Center, Doheny Eye Institute; Department of Ophthalmology, David Geffen School of Medicine, University of California Los Angeles, Los Angeles, CA, USA. FAU - Sadda, Srinivas R AU - Sadda SR AD - Doheny Image Reading Center, Doheny Eye Institute; Department of Ophthalmology, David Geffen School of Medicine, University of California Los Angeles, Los Angeles, CA, USA. ssadda@doheny.org. LA - eng PT - Comparative Study PT - Journal Article DEP - 20200326 PL - Germany TA - Graefes Arch Clin Exp Ophthalmol JT - Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie JID - 8205248 SB - IM MH - Adult MH - Anterior Chamber/diagnostic imaging MH - Anterior Eye Segment/*diagnostic imaging MH - Equipment Design MH - Female MH - Healthy Volunteers MH - Humans MH - Male MH - Middle Aged MH - Prospective Studies MH - Reproducibility of Results MH - Tomography, Optical Coherence/*instrumentation MH - Young Adult OTO - NOTNLM OT - AS-OCTs OT - Agreement OT - Anterior chamber angle OT - Metrics OT - Reproducibility EDAT- 2020/03/28 06:00 MHDA- 2021/04/14 06:00 CRDT- 2020/03/28 06:00 PHST- 2019/11/05 00:00 [received] PHST- 2020/03/13 00:00 [accepted] PHST- 2020/03/07 00:00 [revised] PHST- 2020/03/28 06:00 [pubmed] PHST- 2021/04/14 06:00 [medline] PHST- 2020/03/28 06:00 [entrez] AID - 10.1007/s00417-020-04648-1 [pii] AID - 10.1007/s00417-020-04648-1 [doi] PST - ppublish SO - Graefes Arch Clin Exp Ophthalmol. 2020 Jul;258(7):1475-1481. doi: 10.1007/s00417-020-04648-1. Epub 2020 Mar 26.