PMID- 32378125 OWN - NLM STAT- MEDLINE DCOM- 20210601 LR - 20210601 IS - 2662-4737 (Electronic) IS - 2662-4729 (Linking) VI - 43 IP - 2 DP - 2020 Jun TI - Effect of table height displacement and patient center deviation on size-specific dose estimates calculated from computed tomography localizer radiographs. PG - 665-672 LID - 10.1007/s13246-020-00874-3 [doi] AB - The aim of this study is to evaluate the effect of table height displacement and patient center deviation along the [Formula: see text]-axis on size-specific dose estimate (SSDE) calculations based on computed tomography (CT) localizer radiographs in pediatric and adult abdominal CT examinations. CT localizer radiographs and CT axial images were acquired with table heights of - 5.0, - 2.5, 0.0 (center), 2.5, and 5.0 cm using two acrylic self-made phantoms filled with water. Water-equivalent diameters ([Formula: see text]) were calculated from the CT localizer radiographs and CT axial images. Relative errors of SSDEs from the CT localizer radiographs to SSDEs from the CT axial images were calculated to evaluate the effect of table height displacement. Furthermore, patient center deviations and indices of SSDE overestimation were measured from the clinical data of 110 abdominal CT examinations. The relative errors of SSDEs in phantoms equivalent to 1-year-old and 20-year-old Japanese reference persons ranged from - 2.45% (table height of 50 mm) to + 1.88% (- 50 mm) and from - 4.22% (50 mm) to + 3.79% (- 50 mm), respectively. The largest center deviation in all patients ranged from - 43.1 to 21.5 mm (median: - 14.4 mm). The indices of SSDE overestimation for all patients ranged from - 16.2 to 6.9 mm (median: - 2.2 mm). We found that the effects of table height displacement and patient center deviation along the [Formula: see text]-axis on SSDEs calculated from CT localizer radiographs in pediatric phantoms were smaller compared to adult phantoms. In order to correct these patient center deviations, it is necessary to apply an appropriate correction technique in each section along the [Formula: see text]-axis. FAU - Kuriyama, Kazuki AU - Kuriyama K AUID- ORCID: 0000-0001-6045-3560 AD - Department of Radiology, The Jikei University Kashiwa Hospital, 163-1 Kashiwa-shita, Kashiwa-shi, Chiba, 277-8567, Japan. hjkl.a19@gmail.com. AD - Division of Health Sciences, Department of Quantum Medical Technology, Graduate Course of Medical Science and Technology, Graduate School of Medical Science, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa-shi, Ishikawa, 920-0942, Japan. hjkl.a19@gmail.com. FAU - Matsubara, Kosuke AU - Matsubara K AD - Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa-shi, Ishikawa, 920-0942, Japan. FAU - Hisahara, Shu AU - Hisahara S AD - Department of Radiology, The Jikei University Kashiwa Hospital, 163-1 Kashiwa-shita, Kashiwa-shi, Chiba, 277-8567, Japan. FAU - Nagata, Yukie AU - Nagata Y AD - Department of Radiology, The Jikei University Kashiwa Hospital, 163-1 Kashiwa-shita, Kashiwa-shi, Chiba, 277-8567, Japan. FAU - Nosaka, Rumiko AU - Nosaka R AD - Department of Radiology, The Jikei University Kashiwa Hospital, 163-1 Kashiwa-shita, Kashiwa-shi, Chiba, 277-8567, Japan. FAU - Goto, Rika AU - Goto R AD - Department of Radiology, The Jikei University Kashiwa Hospital, 163-1 Kashiwa-shita, Kashiwa-shi, Chiba, 277-8567, Japan. FAU - Yanano, Nobutaka AU - Yanano N AD - Department of Radiology, The Jikei University Kashiwa Hospital, 163-1 Kashiwa-shita, Kashiwa-shi, Chiba, 277-8567, Japan. FAU - Shimizu, Kanichiro AU - Shimizu K AD - Department of Radiology, The Jikei University School of Medicine, Kashiwa Hospital, 163-1 Kashiwa-shita, Kashiwa-shi, Chiba, 277-8567, Japan. FAU - Shoji, Tomokazu AU - Shoji T AD - Department of Radiology, The Jikei University Hospital, 3-19-18 Nishishinbashi, Minato-ku, Tokyo, 105-8471, Japan. LA - eng GR - JP18K07746/Japan Society for the Promotion of Science/ PT - Journal Article DEP - 20200506 PL - Switzerland TA - Phys Eng Sci Med JT - Physical and engineering sciences in medicine JID - 101760671 SB - IM MH - Humans MH - Infant MH - Phantoms, Imaging MH - *Radiation Dosage MH - *Tomography, X-Ray Computed MH - Young Adult OTO - NOTNLM OT - Computed tomography OT - Localizer radiograph OT - Miscentering OT - Size-specific dose estimate OT - Table height EDAT- 2020/05/08 06:00 MHDA- 2021/06/02 06:00 CRDT- 2020/05/08 06:00 PHST- 2020/02/20 00:00 [received] PHST- 2020/04/22 00:00 [accepted] PHST- 2020/05/08 06:00 [pubmed] PHST- 2021/06/02 06:00 [medline] PHST- 2020/05/08 06:00 [entrez] AID - 10.1007/s13246-020-00874-3 [pii] AID - 10.1007/s13246-020-00874-3 [doi] PST - ppublish SO - Phys Eng Sci Med. 2020 Jun;43(2):665-672. doi: 10.1007/s13246-020-00874-3. Epub 2020 May 6.