PMID- 32585415 OWN - NLM STAT- MEDLINE DCOM- 20201124 LR - 20201124 IS - 1873-4499 (Electronic) IS - 0899-7071 (Linking) VI - 68 DP - 2020 Dec TI - Open 1.0-T versus closed 1.5-T cardiac MR: Image quality assessment. PG - 102-107 LID - S0899-7071(20)30225-4 [pii] LID - 10.1016/j.clinimag.2020.06.019 [doi] AB - PURPOSE: The aim of this paper was to compare the open 1-T (O-1T) versus the closed 1.5-T (C-1.5T) cardiac magnetic resonance (MR). PATIENTS/METHODS: The MR examinations of two concurrent cohorts (each including 100 subjects) of patients with suspected or known cardiac disease were reviewed. Such examinations were obtained using O-1T or C-1.5T MRI. The bright-blood cine, T1-weighted (T1), T2-weighed short-tau inversion recovery (T2-STIR), late gadolinium enhancement (LGE) sequences were performed. Signal-to-noise ratio of blood (SNRb) or myocardium (SNRm), and contrast-to-noise ratio of myocardium (CNRm) were calculated. Subjective image quality (SIQ) of each sequence was graded as 0 = poor, 1 = intermediate, or 2 = optimal. Each examination was considered as diagnostic when the report answered the clinical question. RESULTS: C-1.5T was better than O-1T on cine for SNRb(median 172 versus 452), SNRm(71 versus 160) and CNRm (107 versus 265) and on T2-STIR for SNRb(10 versus 29), SNRm(74 versus 261) and CNRm(-67 versus -233)(P < 0.001). On LGE, SNRm was higher with O-1T than for C-1.5T (312 versus 79, P < 0.001) while CNR was lower (158 versus 389; P < 0.001). No significant differences were found for SNRb on LGE and both SNRm and CNRm on T1 (P >/= 0.215). SIQ of O-1T was not significantly different from that of C-1.5T for both R1 and R2 for cine, T1, and LGE (P >/= 0.157); for T2-STIR, SIQ of O-1T was significantly lower (P = 0.003). R1-R2 concordance was almost perfect (kappa = 0.816-0.894), and all examinations were diagnostic. CONCLUSION: Even though quantitative measurements mostly favored C-1.5T, the SIQ of O-1T was not significantly different for any sequence, with the only exception of T2-STIR. CI - Copyright (c) 2020. Published by Elsevier Inc. FAU - Ali, Marco AU - Ali M AD - Unit of Diagnostic Imaging and Stereotactic Radiosurgery, CDI Centro Diagnostico Italiano, Via Saint Bon 20, 20147 Milan, Italy; Radiology Unit, IRCCS Policlinico San Donato, Via Morandi 30, 20097 San Donato Milanese, Italy. Electronic address: marco.ali@cdi.it. FAU - Monti, Caterina Beatrice AU - Monti CB AD - Department of Biomedical Sciences for Health, Universita degli Studi di Milano, Via Mangiagalli 31, 20133 Milano, Italy. Electronic address: caterina.monti@unimi.it. FAU - Gold, Barak AU - Gold B AD - Corso di Laurea in Medicina e Chirurgia, Universita degli Studi di Milano, Via Festa del Perdono 7, 20122 Milan, Italy. FAU - Lastella, Giulia AU - Lastella G AD - Postgraduation School in Radiodiagnostics, Universita degli Studi di Milano, Via Festa del Perdono 7, 20122 Milano, Italy. FAU - Papa, Sergio AU - Papa S AD - Unit of Diagnostic Imaging and Stereotactic Radiosurgery, CDI Centro Diagnostico Italiano, Via Saint Bon 20, 20147 Milan, Italy. Electronic address: sergio.papa@cdi.it. FAU - Sardanelli, Francesco AU - Sardanelli F AD - Radiology Unit, IRCCS Policlinico San Donato, Via Morandi 30, 20097 San Donato Milanese, Italy; Department of Biomedical Sciences for Health, Universita degli Studi di Milano, Via Morandi 30, 20097 San Donato Milanese, Italy. Electronic address: francesco.sardanelli@unimi.it. FAU - Secchi, Francesco AU - Secchi F AD - Radiology Unit, IRCCS Policlinico San Donato, Via Morandi 30, 20097 San Donato Milanese, Italy; Department of Biomedical Sciences for Health, Universita degli Studi di Milano, Via Morandi 30, 20097 San Donato Milanese, Italy. Electronic address: francesco.secchi@grupposandonato.it. LA - eng PT - Journal Article DEP - 20200615 PL - United States TA - Clin Imaging JT - Clinical imaging JID - 8911831 RN - 0 (Contrast Media) RN - AU0V1LM3JT (Gadolinium) SB - IM MH - *Contrast Media MH - *Gadolinium MH - Heart MH - Humans MH - Magnetic Resonance Imaging MH - Magnetic Resonance Spectroscopy OTO - NOTNLM OT - Cardiac magnetic resonance OT - Claustrophobic patients OT - Obese patients OT - Open magnetic resonance OT - Reproducibility EDAT- 2020/06/26 06:00 MHDA- 2020/11/25 06:00 CRDT- 2020/06/26 06:00 PHST- 2020/02/20 00:00 [received] PHST- 2020/05/13 00:00 [revised] PHST- 2020/06/12 00:00 [accepted] PHST- 2020/06/26 06:00 [pubmed] PHST- 2020/11/25 06:00 [medline] PHST- 2020/06/26 06:00 [entrez] AID - S0899-7071(20)30225-4 [pii] AID - 10.1016/j.clinimag.2020.06.019 [doi] PST - ppublish SO - Clin Imaging. 2020 Dec;68:102-107. doi: 10.1016/j.clinimag.2020.06.019. Epub 2020 Jun 15.