PMID- 33763763 OWN - NLM STAT- MEDLINE DCOM- 20211015 LR - 20211015 IS - 1352-8661 (Electronic) IS - 0968-5243 (Linking) VI - 34 IP - 5 DP - 2021 Oct TI - A simple apparatus for safety assessment of magnetically induced torque on active implantable medical devices (AIMDs) under 1.5 T and 3.0 T MRI. PG - 767-774 LID - 10.1007/s10334-021-00921-4 [doi] AB - OBJECTIVE: Our goal is to design and validate a simple apparatus for the safety assessments of magnetically induced torques by four active implantable medical devices (AIMDs) in 1.5 T and 3.0 T magnetic resonance imaging (MRI) environments. MATERIALS AND METHODS: A simple apparatus was designed to measure the magnetically induced torque. Before testing the apparatus, its effectiveness was first evaluated with a commercially available weights. The torque values of the four AIMDs, namely implantable cardioverter defibrillator (ICD), pacemaker, bone conduction implant, and cochlear implant, under 1.5 T and 3.0 T MRI environments were measured and compared with established international standard. RESULTS: The ICD, pacemaker, bone conduction implant, and cochlear implant had average torques of 1.405, 0.255, 9.460, and 5.490 N.cm under 1.5 T MRI and 2.950, 0.668, 14.618, and 11.484 N.cm under 3.0 T MRI, respectively. DISCUSSION: The apparatus developed in this study was verified to be feasible for measuring magnetically induced torques, with tests on four types of AIMDs under 1.5 T and 3.0 T MRI environments. Although the designed structure is simple, the apparatus was shown to measure torque values of AIMDs under MRI accurately. CI - (c) 2021. European Society for Magnetic Resonance in Medicine and Biology (ESMRMB). FAU - Seo, Hyo-Chang AU - Seo HC AD - Department of Biomedical Engineering, Asan Medical Institute of Convergence Science and Technology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, 05505, Republic of Korea. FAU - Lee, Yongkwan AU - Lee Y AD - Department of Biomedical Engineering, Asan Medical Institute of Convergence Science and Technology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, 05505, Republic of Korea. FAU - Joo, Segyeong AU - Joo S AUID- ORCID: 0000-0002-1681-3845 AD - Department of Biomedical Engineering, Asan Medical Institute of Convergence Science and Technology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, 05505, Republic of Korea. sgjoo@amc.seoul.kr. LA - eng GR - 17172MFDS361/Ministry of Food and Drug Safety, Republic of Korea/ PT - Journal Article DEP - 20210325 PL - Germany TA - MAGMA JT - Magma (New York, N.Y.) JID - 9310752 SB - IM MH - *Cochlear Implants MH - Equipment Safety MH - Magnetic Resonance Imaging MH - *Pacemaker, Artificial MH - Torque OTO - NOTNLM OT - Compliance standards OT - Magnetic resonance imaging OT - Medical device safety OT - Torque EDAT- 2021/03/26 06:00 MHDA- 2021/10/16 06:00 CRDT- 2021/03/25 06:40 PHST- 2020/11/13 00:00 [received] PHST- 2021/03/11 00:00 [accepted] PHST- 2021/03/10 00:00 [revised] PHST- 2021/03/26 06:00 [pubmed] PHST- 2021/10/16 06:00 [medline] PHST- 2021/03/25 06:40 [entrez] AID - 10.1007/s10334-021-00921-4 [pii] AID - 10.1007/s10334-021-00921-4 [doi] PST - ppublish SO - MAGMA. 2021 Oct;34(5):767-774. doi: 10.1007/s10334-021-00921-4. Epub 2021 Mar 25.