PMID- 34292121 OWN - NLM STAT- MEDLINE DCOM- 20211015 LR - 20240403 IS - 1943-4936 (Electronic) IS - 1040-6387 (Print) IS - 1040-6387 (Linking) VI - 33 IP - 5 DP - 2021 Sep TI - Validation study of canine serum cortisol measurement with the Immulite 2000 Xpi cortisol immunoassay. PG - 844-863 LID - 10.1177/10406387211029247 [doi] AB - We report the results of validation of canine serum cortisol determination with the Immulite 2000 Xpi cortisol immunoassay (Siemens), with characterization of precision (CV), accuracy (spiking-recovery [SR] bias), and observed total error (TEo = bias + 2CV) across the reportable range, specifically at the most common interpretation thresholds for dynamic testing. Imprecision increased at increasing rate with decreasing serum cortisol concentration and bias was low, resulting in increasing TEo with decreasing serum cortisol concentration. Inter-laboratory comparison study allowed for determination of range-based bias (RB) and average bias (AB). At 38.6 and 552 nmol/L (1.4 and 20 mug/dL), between-run CV was 10% and 7.5%, respectively, and TEo ~30% and ~20%, respectively (TEo remained similar regardless of the considered bias: SR, RB, or AB). These analytical performance parameters should be considered in the interpretation of results and for future expert consensus discussions to determine recommendations for allowable total error (TEa). Importantly, the commonly used thresholds for interpretation of results were determined ~40 y ago with different methods of measurements and computation, hence updating is desirable. Quality control material (QCM) had between-run imprecision of 4% for QCM1 and 7% for QCM2; the bias was minimal for both levels. Acceptable QC rules are heavily dependent on the desired TEa for the QCM system (TEa(QCM)), itself limited by the desired clinical TEa. At low TEa(QCM) (20-33%), almost no rules were acceptable, whereas at high TEa(QCM) (50%), almost all rules were acceptable; further investigation is needed to determine which TEa(QCM) can be guaranteed by simple QC rules. FAU - Korchia, Jeremie AU - Korchia J AUID- ORCID: 0000-0002-9344-6639 AD - Texas A&M Veterinary Medical Diagnostic Laboratory, Texas A&M University, College Station, TX, USA. FAU - Freeman, Kathleen P AU - Freeman KP AUID- ORCID: 0000-0003-1796-0158 AD - SYNLAB-VPG/Exeter, Exeter, United Kingdom. LA - eng PT - Journal Article DEP - 20210722 PL - United States TA - J Vet Diagn Invest JT - Journal of veterinary diagnostic investigation : official publication of the American Association of Veterinary Laboratory Diagnosticians, Inc JID - 9011490 RN - WI4X0X7BPJ (Hydrocortisone) SB - IM MH - Animals MH - Dogs MH - *Hydrocortisone MH - Immunoassay/veterinary PMC - PMC8366243 OTO - NOTNLM OT - Cushing OT - accuracy OT - bias OT - cortisol OT - dogs OT - endocrinology OT - precision OT - total error COIS- Declaration of conflicting interests: The authors declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article. EDAT- 2021/07/23 06:00 MHDA- 2021/10/16 06:00 PMCR- 2022/07/22 CRDT- 2021/07/22 12:19 PHST- 2021/07/23 06:00 [pubmed] PHST- 2021/10/16 06:00 [medline] PHST- 2021/07/22 12:19 [entrez] PHST- 2022/07/22 00:00 [pmc-release] AID - 10.1177_10406387211029247 [pii] AID - 10.1177/10406387211029247 [doi] PST - ppublish SO - J Vet Diagn Invest. 2021 Sep;33(5):844-863. doi: 10.1177/10406387211029247. Epub 2021 Jul 22.