PMID- 34657257 OWN - NLM STAT- MEDLINE DCOM- 20220301 LR - 20220301 IS - 1614-7499 (Electronic) IS - 0944-1344 (Linking) VI - 29 IP - 12 DP - 2022 Mar TI - Fully-automated SPE coupled to UHPLC-MS/MS method for multiresidue analysis of 26 trace antibiotics in environmental waters: SPE optimization and method validation. PG - 16973-16987 LID - 10.1007/s11356-021-15947-3 [doi] AB - Achieving simultaneous determination of antibiotic multiresidues in environmental waters by solid phase extraction (SPE) coupled with ultra-high-performance liquid chromatography tandem mass spectrometry (UHPLC-MS/MS) with detection limits 0.995) and precision (< 9.9%). Recoveries in spike samples ranged between 77.5-104.9% in pure water, 59.4-97.8% in surface water (SW), and 58.2-108.6% in wastewater effluent (WWE) with relative standard deviations