PMID- 35040673 OWN - NLM STAT- MEDLINE DCOM- 20220427 LR - 20220524 IS - 1527-1315 (Electronic) IS - 0033-8419 (Print) IS - 0033-8419 (Linking) VI - 303 IP - 2 DP - 2022 May TI - Dependence of Water-equivalent Diameter and Size-specific Dose Estimates on CT Tube Potential. PG - 404-411 LID - 10.1148/radiol.210860 [doi] AB - Background The size-specific dose estimate (SSDE) is a patient-focused CT dose metric. However, published size-dependent conversion factors (f(size)) used to calculate SSDE were determined primarily by using phantoms; only eight to 15 patient data sets were used, all at 120 kV. Purpose To determine the effect of different tube potentials on the water-equivalent diameter (WED) and SSDE for patient CT scans of the head, chest, and abdomen. Materials and Methods This retrospective study used 250 noncontrast CT scans acquired between March 2013 and June 2017. Bony structures were segmented, and their CT numbers were modified to reflect bone attenuation at 70, 90, 110, 130, and 150 kV. Soft-tissue CT numbers were unchanged because of negligible energy dependence. f(size) was measured in anthropomorphic phantoms for each tube potential and fit to an exponential function. WED and SSDE were determined for each patient at all tube potentials, regression analysis was performed relative to the WED and SSDE at 120 kV, and mean differences relative to 120 kV were calculated. Results In 250 patients (median age, 21.5 years; interquartile range, 44 years; 130 women), WED for all tube potentials was linearly related to the WED at 120 kV in all body regions (R(2) = 0.995-1.000). The effect of tube potential on WED was negligible for torso examinations (Cohen d < 0.05). In the head, a medium effect size was observed at 70 kV; however, the mean absolute difference in WED was small (-0.49 cm +/- 0.08 [standard deviation]; P < .001). For commonly used combinations of tube potential and patient size, the mean differences in SSDE at alternative tube potentials relative to SSDE at 120 kV were less than 5%. Conclusion At noncontrast CT, published size-dependent conversion factors accurately determined size-specific dose estimates on 250 patient scans at five tube potentials other than 120 kV. (c) RSNA, 2022 Online supplemental material is available for this article. See also the editorial by Boone in this issue. FAU - McCollough, Cynthia H AU - McCollough CH AUID- ORCID: 0000-0002-5346-332X AD - From the Department of Radiology, Mayo Clinic, 200 First St SW, Rochester, MN 55905. FAU - McCollough, Shannon L AU - McCollough SL AUID- ORCID: 0000-0002-2768-0435 AD - From the Department of Radiology, Mayo Clinic, 200 First St SW, Rochester, MN 55905. FAU - Schneider, Justine J AU - Schneider JJ AUID- ORCID: 0000-0002-3519-4752 AD - From the Department of Radiology, Mayo Clinic, 200 First St SW, Rochester, MN 55905. FAU - Moen, Taylor R AU - Moen TR AD - From the Department of Radiology, Mayo Clinic, 200 First St SW, Rochester, MN 55905. FAU - Weaver, Jayse M AU - Weaver JM AUID- ORCID: 0000-0002-6123-8878 AD - From the Department of Radiology, Mayo Clinic, 200 First St SW, Rochester, MN 55905. FAU - Vrieze, Thomas J AU - Vrieze TJ AD - From the Department of Radiology, Mayo Clinic, 200 First St SW, Rochester, MN 55905. FAU - Yu, Lifeng AU - Yu L AUID- ORCID: 0000-0002-4267-1326 AD - From the Department of Radiology, Mayo Clinic, 200 First St SW, Rochester, MN 55905. FAU - Leng, Shuai AU - Leng S AUID- ORCID: 0000-0002-6453-9481 AD - From the Department of Radiology, Mayo Clinic, 200 First St SW, Rochester, MN 55905. LA - eng GR - R01 EB017095/EB/NIBIB NIH HHS/United States PT - Journal Article PT - Research Support, N.I.H., Extramural PT - Research Support, Non-U.S. Gov't DEP - 20220118 PL - United States TA - Radiology JT - Radiology JID - 0401260 RN - 059QF0KO0R (Water) SB - IM CIN - Radiology. 2022 Jan 18;:212801. PMID: 35040681 MH - Adult MH - Female MH - Humans MH - Male MH - Phantoms, Imaging MH - Radiation Dosage MH - Retrospective Studies MH - *Tomography, X-Ray Computed/methods MH - *Water MH - Young Adult PMC - PMC9038617 MID - NIHMS1755829 COIS- Disclosures of Conflicts of Interest: C.H.M. Travel support from American Association of Physicists in Medicine (AAPM); president and chair of the Board of the AAPM; vice-president and board member of International Society of CT. S.L.M. No relevant relationships. J.J.S. No relevant relationships. T.R.M. No relevant relationships. J.M.W. No relevant relationships. T.J.V. No relevant relationships. L.Y. No relevant relationships. S.L. No relevant relationships. EDAT- 2022/01/19 06:00 MHDA- 2022/04/28 06:00 PMCR- 2023/05/01 CRDT- 2022/01/18 12:19 PHST- 2022/01/19 06:00 [pubmed] PHST- 2022/04/28 06:00 [medline] PHST- 2022/01/18 12:19 [entrez] PHST- 2023/05/01 00:00 [pmc-release] AID - 10.1148/radiol.210860 [doi] PST - ppublish SO - Radiology. 2022 May;303(2):404-411. doi: 10.1148/radiol.210860. Epub 2022 Jan 18.