PMID- 35799986 OWN - NLM STAT- MEDLINE DCOM- 20220711 LR - 20220716 IS - 1846-7482 (Electronic) IS - 1330-0962 (Print) IS - 1330-0962 (Linking) VI - 32 IP - 2 DP - 2022 Jun 15 TI - Performance evaluation of the high sensitive troponin I assay on the Atellica IM analyser. PG - 020709 LID - 10.11613/BM.2022.020709 [doi] LID - 020709 AB - INTRODUCTION: The Fourth Universal Definition of Myocardial Infarction Global Taskforce recommends the use of high sensitive troponin (hs-Tn) assays in the diagnosis of acute myocardial infarction. We evaluated the analytical performance of the Atellica IM High-sensitivity Troponin I Assay (hs-TnI) (Siemens Healthcare Diagnostics Inc., Tarrytown, USA) and compared its performance to other hs-TnI assays (Siemens Advia Centaur, Dimension Vista, Dimension EXL, and Abbott Architect (Wiesbaden, Germany)) at one or more sites across Europe. MATERIALS AND METHODS: Precision, detection limit, linearity, method comparison, and interference studies were performed according to Clinical and Laboratory Standards Institute protocols. Values in 40 healthy individuals were compared to the manufacturer's cut-offs. Sample turnaround time (TAT) was examined. RESULTS: Imprecision repeatability CVs were 1.1-4.7% and within-lab imprecision were 1.8-7.6% (10.0-25,000 ng/L). The limit of blank (LoB), detection (LoD), and quantitation (LoQ) aligned with the manufacturer's values of 0.5 ng/L, 1.6 ng/L, and 2.5 ng/L, respectively. Passing-Bablok regression demonstrated good correlations between Atellica IM analyser with other systems; some minor deviations were observed. All results in healthy volunteers fell below the 99th percentile URL, and greater than 50% of each sex demonstrated values above the LoD. No interference was observed for biotin (