PMID- 38335725 OWN - NLM STAT- MEDLINE DCOM- 20240403 LR - 20240403 IS - 1943-5681 (Electronic) IS - 0002-9645 (Linking) VI - 85 IP - 4 DP - 2024 Apr 1 TI - The Edwards Acumen IQ system using peripheral arterial catheter-based waveforms to estimate cardiac output is not accurate as compared to thermodilution in dogs. LID - ajvr.23.11.0249 [pii] LID - 10.2460/ajvr.23.11.0249 [doi] AB - OBJECTIVE: To assess the agreement between cardiac output (CO) estimated via evaluation of the arterial pressure waveform by a novel monitoring system (Edwards Acumen IQ sensor and HemoSphere Advanced Monitor Platform [HS-IQ]; Edwards LifeSciences) and measured by thermodilution (TD) in anesthetized, normovolemic, and hypovolemic dogs. To assess the agreement between the HS-IQ CO measurements in the radial artery and dorsal metatarsal artery. ANIMALS: 8 purpose-bred Beagles. METHODS: Dogs were placed under general anesthesia. CO was measured via TD and via the HS-IQ at radial and dorsal metatarsal arterial catheters. CO measurements were obtained at 4 time points including normovolemic and multiple hypovolemic states. Paired measurements of CO were evaluated via the method of Bland and Altman with acceptable limits of agreement (LOA) defined as < 30%. RESULTS: A total of 24 (dorsal metatarsal) and 21 (radial) paired measurements were collected in 8 dogs. The overall bias (CI) for comparison of TD to radial arterial HS-IQ CO measurements was -0.09 L/min. LOA and proportional LOA were -2.66 to 2.49 L/min and -140.72% to 104.94%. The overall bias (CI) for comparison of TD to dorsal metatarsal arterial HS-IQ CO measurements was -0.26 L/min. LOA and proportional LOA were -2.76 to 2.24 L/min and -135.96% to 93.25%. The overall proportional error for radial arterial was -17.9% and for dorsal metatarsal was -21.4%. CLINICAL RELEVANCE: CO measurements with the HS-IQ were easy to obtain but did not produce results within a clinically acceptable range for either measurement site, with a very wide LOA. The CO estimations from the HS-IQ are not appropriate for clinical use at this time. FAU - Zersen, Kristin M AU - Zersen KM AUID- ORCID: 0000-0003-0327-8537 FAU - Griffenhagen, Gregg M AU - Griffenhagen GM AUID- ORCID: 0000-0003-3522-5587 FAU - Monnet, Eric L AU - Monnet EL AUID- ORCID: 0000-0002-0058-2210 LA - eng PT - Journal Article DEP - 20240213 PL - United States TA - Am J Vet Res JT - American journal of veterinary research JID - 0375011 SB - IM MH - Dogs MH - Animals MH - *Thermodilution/veterinary/methods MH - Hypovolemia/veterinary MH - Monitoring, Intraoperative/methods/veterinary MH - Cardiac Output MH - Arteries MH - Catheters, Indwelling MH - Reproducibility of Results MH - *Dog Diseases/diagnosis OTO - NOTNLM OT - arterial catheter OT - cardiac output OT - hypovolemia OT - normovolemia OT - thermodilution EDAT- 2024/02/10 13:18 MHDA- 2024/04/03 06:44 CRDT- 2024/02/09 18:06 PHST- 2023/11/07 00:00 [received] PHST- 2024/01/24 00:00 [accepted] PHST- 2024/04/03 06:44 [medline] PHST- 2024/02/10 13:18 [pubmed] PHST- 2024/02/09 18:06 [entrez] AID - ajvr.23.11.0249 [pii] AID - 10.2460/ajvr.23.11.0249 [doi] PST - epublish SO - Am J Vet Res. 2024 Feb 13;85(4):ajvr.23.11.0249. doi: 10.2460/ajvr.23.11.0249. Print 2024 Apr 1.