PMID- 8167281 OWN - NLM STAT- MEDLINE DCOM- 19940601 LR - 20220318 IS - 1053-0770 (Print) IS - 1053-0770 (Linking) VI - 8 IP - 1 DP - 1994 Feb TI - Intraoperative cardiac output monitoring: comparison of impedance cardiography and thermodilution. PG - 24-9 AB - Impedance cardiography (IC) is a noninvasive, simple to use method of cardiac output (CO) determination. A prospective evaluation of IC monitoring was performed in 50 patients undergoing noncardiac surgery. IC CO measurements (NC-COM3-Revision 7, BoMed Manufacturing) were compared to simultaneous measurements of thermodilution (TD) CO to assess the validity of this technique for intraoperative cardiac monitoring. Adequate impedance signals could not be obtained in 7 of the 50 patients. IC CO measurements were highly correlated to TD CO (P < .005), with a correlation coefficient r = 0.84. Bias analysis, however, indicated clinically significant disagreement between the two techniques. IC CO tended to underestimate TD CO (mean bias = -0.41 L/min) and the SD of the bias was 1.0 L/min (95% level of agreement 1.6 to -2.4 L/min). Trending data showed IC to accurately track the direction of TD CO changes but to underestimate their magnitude (r = 0.60, intercept -0.7 L/min, slope 0.47). Factors that may have impaired the performance of IC in this study include the high prevalence of cardiac disease in the study population and electrical noise in the operative setting. Further development of IC appears warranted if it is to prove useful as an intraoperative cardiac monitor. FAU - Perrino, A C Jr AU - Perrino AC Jr AD - Department of Anesthesiology, Yale University School of Medicine, New Haven, CT 06510. FAU - Lippman, A AU - Lippman A FAU - Ariyan, C AU - Ariyan C FAU - O'Connor, T Z AU - O'Connor TZ FAU - Luther, M AU - Luther M LA - eng PT - Comparative Study PT - Journal Article PL - United States TA - J Cardiothorac Vasc Anesth JT - Journal of cardiothoracic and vascular anesthesia JID - 9110208 SB - IM MH - Aged MH - Bias MH - *Cardiac Output MH - *Cardiography, Impedance/instrumentation/statistics & numerical data MH - Catheterization MH - Coronary Disease/physiopathology MH - Electrocardiography MH - Humans MH - *Monitoring, Intraoperative/instrumentation MH - Prospective Studies MH - Pulmonary Artery MH - ROC Curve MH - Reproducibility of Results MH - Sensitivity and Specificity MH - Signal Processing, Computer-Assisted MH - Stroke Volume MH - *Thermodilution/instrumentation/statistics & numerical data EDAT- 1994/02/01 00:00 MHDA- 2001/03/28 10:01 CRDT- 1994/02/01 00:00 PHST- 1994/02/01 00:00 [pubmed] PHST- 2001/03/28 10:01 [medline] PHST- 1994/02/01 00:00 [entrez] AID - 1053-0770(94)90007-8 [pii] AID - 10.1016/1053-0770(94)90007-8 [doi] PST - ppublish SO - J Cardiothorac Vasc Anesth. 1994 Feb;8(1):24-9. doi: 10.1016/1053-0770(94)90007-8.