PMID- 8841858 OWN - NLM STAT- MEDLINE DCOM- 19961212 LR - 20171206 IS - 0391-3988 (Print) IS - 0391-3988 (Linking) VI - 19 IP - 7 DP - 1996 Jul TI - The Fresenius continuous autotransfusion system (CATS): preliminary studies and application. PG - 431-4 AB - A new CFC apparatus for intraoperative blood salvage (CATS) has been introduced by Fresenius and submitted to preliminary evaluation at our hemapheresis unit. The volume of shed blood submitted to washing was 1714 +/- 496 with a hematocrit of 19% and the volume of PRBC ready for transfusion 297.9 ml with hematocrit of 64.3%. Elimination of contaminant activated platelets and WBC was 92% and 74% respectively. Elimination of contaminants such as protein, free hemoglobin, LDH and K ions was always over 95%. Regular application of CATS is warrented after our study. FAU - Florio, G AU - Florio G AD - Immunohematology and Transfusion Services, San Martino University Hospital, Genova, Italy. FAU - Valbonesi, M AU - Valbonesi M FAU - Lercari, G AU - Lercari G FAU - Frisoni, R AU - Frisoni R FAU - Pollicardo, N AU - Pollicardo N FAU - Beraudo, S AU - Beraudo S LA - eng PT - Journal Article PL - United States TA - Int J Artif Organs JT - The International journal of artificial organs JID - 7802649 SB - IM MH - *Blood Transfusion, Autologous MH - Cell Size/physiology MH - Erythrocytes/*cytology MH - Hematocrit MH - Humans MH - Intraoperative Complications MH - Leukocytes/*cytology MH - Platelet Activation MH - Retrospective Studies EDAT- 1996/07/01 00:00 MHDA- 1996/07/01 00:01 CRDT- 1996/07/01 00:00 PHST- 1996/07/01 00:00 [pubmed] PHST- 1996/07/01 00:01 [medline] PHST- 1996/07/01 00:00 [entrez] PST - ppublish SO - Int J Artif Organs. 1996 Jul;19(7):431-4.