PMID- 9084523 OWN - NLM STAT- MEDLINE DCOM- 19970417 LR - 20150901 IS - 0254-1319 (Print) IS - 0254-1319 (Linking) VI - 34 IP - 2 DP - 1996 Jun TI - The accuracy of bioimpedance cardiography in the measurement of cardiac output in comparison with thermodilution method. PG - 55-9 AB - BACKGROUND: Continuous noninvasive measurement of cardiac output (CO) is clinically desirable. Thoracic electrical bioimpedance (TEB) offers such a continuous noninvasive technique for the measurement of CO on a beat-by-beat basis. The purpose of this study was to compare the correlation of TEB with the thermodilution (TD) method of measuring CO during surgery. METHODS: Eighteen patients (age 39-64 yr), with coronary artery disease undergoing coronary artery bypass graft surgery were studied. Patients were excluded if severe arrhythmias or aortic insufficiency was present before operation. Simultaneous measurements of CO were performed in each patient with the use of a 7F Swan-Ganz catheter and TEB. 128 matched pair values were obtained and compared using regression analysis. Bias (mean difference between measurements) and limits of agreement with 95% confidence limits were also calculated. RESULTS: A good correlation (correlation coefficient, r = 0.86; TEB = 0.55 + 0.82TD, p < 0.001) between the two methods of measuring CO was observed, and the standard error was small (SE = 0.218). The mean difference between CO measurements (TEB-TD) was -0.66 L/min and had a 95% confidence interval of 0.12 to 0.98 L/min. The standard deviation of the difference was 0.915 L/min and the limits of agreement were from -2.49 to 1.17 L/min. There appeared to be a good agreement between TEB and TD CO measurement in this study. CONCLUSIONS: Technological advances in TEB have resulted in more accurate CO measurements. Such device offers a noninvasive alternative to the TD technique for the determination of CO, and may potentially be very useful as a powerful, noninvasive, continuous technique for monitoring cardiac function in the clinical setting. FAU - Wong, K L AU - Wong KL AD - Department of Anesthesia, Mackay Memorial Hospital, Taipei, Taiwan, R.O.C. FAU - Hou, P C AU - Hou PC LA - eng PT - Comparative Study PT - Journal Article PL - China (Republic : 1949- ) TA - Acta Anaesthesiol Sin JT - Acta anaesthesiologica Sinica JID - 9432542 SB - IM EIN - Acta Anaesthesiol Sin 1996 Sep;34(3):172 MH - Adult MH - *Cardiac Output MH - Electric Impedance MH - Female MH - Humans MH - Male MH - Middle Aged MH - *Thermodilution EDAT- 1996/06/01 00:00 MHDA- 1996/06/01 00:01 CRDT- 1996/06/01 00:00 PHST- 1996/06/01 00:00 [pubmed] PHST- 1996/06/01 00:01 [medline] PHST- 1996/06/01 00:00 [entrez] PST - ppublish SO - Acta Anaesthesiol Sin. 1996 Jun;34(2):55-9.