PMID- 9141314 OWN - NLM STAT- MEDLINE DCOM- 19970522 LR - 20161021 IS - 0158-9938 (Print) IS - 0158-9938 (Linking) VI - 20 IP - 1 DP - 1997 Mar TI - Field programmable gate array in a fast 256-channel data acquisition system. PG - 47-52 AB - We describe an FPGA (Field Programmable Gate Array) controlled fast data acquisition system with 256 channels suitable for high resolution electrocardiographic (ECG), electrogastrographic (EGG) and late potential (LP) recordings. The A/D converter provides 800ns conversion time with 12 bits. The S11W, a 16 bit, parallel, I/O interface, is used to connect the data acquisition device to a portable SPARC-compatible computer. With only one A/D converter, we have achieved 2 kHz sampling rate for each channel and more than 1MB/s overall by using an overlapping delay technique. Channel parameters can be selected independently from one of eight gains ranging from 50 to 10,000, and one of three filters ("ECG" 0.05-120Hz, "EGG" 0.001-1Hz and "LP" 60-1kHz). There are eight programmable sampling frequencies from 2Hz to 2kHz. The system is fully battery operated and protected from the defibrillation high voltages to meet the safety requirements. The acquisition system can be widely used in medical research and clinical applications. FAU - Li, C Y AU - Li CY AD - Div. of Clinical Sciences, Univ. of Tasmania, Hobart, Australia. FAU - Pluta, W K AU - Pluta WK FAU - Kilpatrick, D AU - Kilpatrick D LA - eng PT - Journal Article PL - Netherlands TA - Australas Phys Eng Sci Med JT - Australasian physical & engineering sciences in medicine JID - 8208130 SB - IM MH - Analog-Digital Conversion MH - Electric Impedance MH - Electric Power Supplies MH - Electricity MH - Electrocardiography MH - Evoked Potentials MH - *Signal Processing, Computer-Assisted EDAT- 1997/03/01 00:00 MHDA- 1997/03/01 00:01 CRDT- 1997/03/01 00:00 PHST- 1997/03/01 00:00 [pubmed] PHST- 1997/03/01 00:01 [medline] PHST- 1997/03/01 00:00 [entrez] PST - ppublish SO - Australas Phys Eng Sci Med. 1997 Mar;20(1):47-52.